Samples arc transported upright to the laboratory in a rigid container with packing
material to prevent jostling or damage. To avoid electrostatic forces from causing fiber
loss from the sampling filter, untreated polystyrene foam will not be used for any of the
shipping container materials. Samples for asbestos will be collect using two different
media to accommodate the SEM and TEM analyses. Equipment and supplies utilized
with this sampling approach are as follows:
• Area Sampling Pump - 5 to 15 1pm.
• Filters/TEM analysis - Mixed cellulose ester membrane filters, 0.45-|Jm pore
size, with a non-conductive cowl, supported by a cassette filter holder, suitable for
connection with the personal monitoring pump system.
• Filters/SEM analysis -Polycarbonate filters, 01 .pm pore size with 25 mm
diameter cassette.
• High Flow Rotometer - 2.0 to 20.0 1pm
10.2.4 Respirable Crystalline Silica and Dust
Respirable dust and crystalline sampling will be performed according to NIOSH Method
0600 - Particulates Not Otherwise Regulated, Respirable following the
recommendations for personal sampling. Samples will be collected at a rate of
approximately two liters per minute (2 1pm) for a minimum of eight (8) hours for a
resulting total air volume of approximately 1.0 m3. To ensure optimum accuracy, the
sampling rate will be adjusted, as needed, not to exceed 2 mg dust loading on the
filter. Filters are pre-weighed to a constant weight in the controlled weighing area. If
static electricity is evident (e.g., filter not releasing easily from the forceps or it attracts
the balance pan), the filter should be passed over an anti-static radiation source. Filters
are assembled in the filter cassettes and the cassettes closed firmly by placing a plug at
each opening of the cassette to prevent leakage around the filter. For at least 2 hours prior
to sampling, the filters are equilibrated in either an environmentally controlled weighing
area or chamber. Pump flow rates of each personal sampling pump, with a representative
sampler in line and cyclone grit pot in place, will be measured before and after sample
collection. The pump will be run for 10 minutes prior to checking the flow rate. A
suitable rotometer (or bubble meter) will be connected to the filter inlet, the flow screw
adjusted to the desired sampling value, and the flow rate observed on the rotometer (or
bubble meter) will be verified as constant for a minimum of 20 seconds. Flow rates will
be recorded on Field Sampling Data Sheets or an equivalent field logbook. The cyclone’s
grit cap is removed prior to use and the interior inspected. If the inside is visibly scored,
replace the cyclone because the dust separation characteristics may be altered. Clean
the interior of the cyclone to prevent re-entrainment of large particles. Assemble the
sampler head checking for alignment of the filter hold er and cyclone in the sampling head
to prevent leakage. Care must be taken to prevent the sampler assembly from inverting at
any time. Immediately prior to collecting a sample, the sampler will be fastened to a
location (i.e., on a tripod) near the individual’s breathing zone, the top cover of the filter
cassette cowl extension (“open face”) will be removed and positioned “face down” in the
sampler. The joint between the extender and monitor body will be wrapped with tape to