g. Ability to collect x-ray maps or particle analysis software (preferably both)

8.0 Reagents 1. Isopropanol, reagent grade [CAS No. 67-63-0] 2. 1.55 or 1.605 Refractive Index Oil

9.0 Sample Storage

Dust samples will be stored in an air-tight container, such as a sealed glass jar. Samples placed in reagents will be labeled appropriately and stored according to laboratory safety standards. Samples prepared for analyses will be stored in a protective container, such as a plastic case or covered petri dish, to prevent contamination.

10.0 Quality Control Quality control is implemented by thoroughly cleaning glassware and spatulas, keeping working surfaces clean, and preventing cross contamination. During ashing, particles may be suspended if slow heating is not achieved. Following the ashing program as outlined will minimize flashing, which can cause particles to become airborne. Covered crucibles will be used to prevent contamination caused by flashing. Used Eppendorf pipette tips and weighing papers will be discarded and new tips and papers will be used for each sample.

Duplicate samples shall be prepared to determine the precision of the analysis. In addition, sample blanks shall be prepared. These blanks are checks for cross contamination during handling of the samples. Blanks shall be prepared at the same time and in the same manner as samples.

10.1 Calibration

Calibration of the EDS system must be completed at least once at the beginning and again at the end of each analytical session. Backscattered electron image (BEI) calibration should be performed at the beginning of the session and anytime the backscattered image brightness and/or contrast is adjusted.

EDS calibration for both qualitative and quantitative (not required by this method but could be useful for identification of particle type) analysis is accomplished by the analysis of a polished carbon-coated reference standard. The recommended material is USGS BIRl-G basalt glass mounted in epoxy in a brass tube, polished, and carbon coated using a carbon evaporator (2, 3).

The calibration reference material should be analyzed at the same operating conditions to be used for the analysis including beam current, accelerating voltage, working distance, detector dead time, and sample tilt (= 0°). For BIRl-G

39

CITY_DEP_WORBY_CANNATA_SUBP_000057577

NYC-WTC_000423050

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000423050Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Buildings in this record

Machine-extracted building matches; verify each source. Buildings, never households.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →