METHODOLOGY (Cent.)
Sample Extraction
Each PUF cartridge was spiked with isotopically-labeled PCDD/PCDF internal standards (Table 1),
combined with the filter, and extracted for 18 hours with toluene in a Soxhlet extractor. The
extract was quantitatively transferred to a Kuderna Danish concentrator, concentrated, solvent
/exchanged to hexane, and processed through the analyte enrichment procedures described below.
Analyte Enrichment for PCDD/PCDF Analyses
The extraction procedure often removes a variety of compounds, in addition to the PCDDs and
PCDFs, from the sample matrix. Some of these compounds can directly interfere with the analyses
while others can overload the capillary column causing degradation in chromatographic resolution
or sensitivity. The analyte enrichment steps described below were used to remove interferences
from the extracts.
Each sample extract was diluted to 100 mL with hexane, transferred to a separatory funnel, and
. washed with IN sodium hydroxide, concentrated sulfuric acid, and aqueous sodium chloride (5%
w/v) as needed. The hexane layer was concentrated to 1 mL and quantitatively transferred to a
liquid chromatography column containing alternating layers of silica gel, 40% concentrated sulfuric
acid on silica gel, and 33% IN sodium hydroxide on silica gel. The column was eluted with 90 mL
of hexane and the entire eluate was collected and concentrated, under ambient conditions, to a
volume of 1 mL.
f Each extract was then fractionated on a liquid chromatography column containing 4 g of activated
alumina. The column was eluted'with 20 mL of hexane followed by 1 5 mL of 60% methylene
chloride/hexane. The 60% methylene chloride/hexane fraction was concentrated to 1 mL under a
stream of dry nitrogen and applied to the top of a chromatography column containing 1 g of 5%
AX-21 activated carbon in silica gel. The column was eluted with two 2 mL portions of hexane, 2
mL of cyclohexane/methylene chloride (50:50 v/v) and cyclohexane/methanol/toluene (75:20:5
v/v) in the forward direction, and then with 20 mL of toluene in the reverse direction. The toluene
fraction was collected, spiked with recovery standards (1,2,3,4-TCDD-'2Ci2 and 1,2,3,7,8,9-
P; HxCDD-’^Ci2) and taken to a final volume of 20 uL.