I [Nancy Jeffery - wtc Silica SOP MSD0700.doc

Material Science Division Operating Procedures for Silica Analysis: MSD 0700 Jan2000

the sample. (See Analytical Curve 9.0}.

9.0 Analytical Curve

The Analytical Curve is a series of reference points for calculation of silica mass in the sample filter. Silica weight standards are prepared by suspending known weights of SRM’s in 2-propanol. The suspension is placed into an ultrasonic bath for fifteen minutes prior to mechanical stirring. Aliquots are filtered through 0.45um silver membrane filters. These weight standards are analyzed following the criteria in section 7.0 to determine the net background subtracted peak intensity. These intensities are graphed and the data points should form a linear correlation based upon weight. Linear regression is used to determine the weight of silica in a sample according to its peak intensity.

10.0 Calibrations:

10.1 The slope of the Analytical Curve should not display an intercept greater than +5ug of zero. 10.2 Repeatability of each standard greater than 10% (in standards >0.04mg) indicates that new standards are required. 10.3 Six unused filters from the same filter lot are subjected to a “Blank” preparation of 2-propanol and analyzed to test for self-absorption by determining the net normalized count for the silver peak for each media batch.

11.0 Calculations:

I, X f(t) - b C= ---------------- M X V

Where: lx = Normalized intensity for sample peak b= intercept of calibration graph m= slope of calibration graph f(t) = -R in T/(1 -T”) = absorption correction factor (NIOSH 7500 Table #1) R= sin (©Ag) /sin (Ox) T= lAg /(average Ug) = transmittance of sample Iao = normalized silver peak 1% = normalized silver peak intensity from media blanks

6

CITY_21MC100_010220830

NYC-WTC_000338491

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000338491Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →