. c,»a. c_«jox O-t-friT NO.432 P.9/14

________ ELEMENTS (ICP1: METHOD 7300, ISSUE 2, dated 1S August 1994 • Page 8 of 9

APPENDIX. MICROWAVE DIGESTION FOR LEAD IN PAINT CHIPS (AND OTHER MATRICES) This procedure is an alternative to the procedure presented In the Sample Preparation section of this method. It provides a rapid, complete acid digestion prior to analysis by flame atomic absorption (FAA), heated graphite furnace atomic absorption (HGFAA), and inductively coupled plasma spectroscopy (ICP) (10]. Apparatus and Materialfl 1-16]

1. Microwave apparatus requirements: a. The microwave unit provides programmable power with a minimum of 574 W and can be programmed to within ± 10 W of the required power. b. The microwave unit cavity is corrosion resistant as well as ventilated. All electronics are protected against corrosion tor safe operation. c. The system requires Teflon PFA digestion vessels (120-mL capacity] capable of withstanding pressures up to 7.5 ± 0.7 atm (110 ± 10 psi) and capable of controlled pressure relief at pressures exceeding 7.610.7 atm (110110 psi), d. A rotating turntable is employed to ensure homogeneous distribution of microwave radiation within the unit. The speed of the turntable should be a minimum of 3 rpm. e, A safety concern relates to the use of sealed containers without pressure relief valves In the unit. Temperatura Is the Important variable controlling (he reaction. Pressure Is needed to attain ele­ vated temperatures but must be safely contained [12]. f. Palymerlc volumetric ware in plastic (Teflon or pc yethylene), 50- or 100-mL capacity. g. Dispoeable polypropylene filler funnel. h. Analytical balance, 300-g capacity, and minimum ± 0.001 g.

Raaqants 1, Nitric acid, concentrated, spectroscopy grade. 2. Reagent Water. Reagent water shall be interference free. All references to water in the method refer to reagent water that meets the ASTM Type 2 standard.

Procedure

1. Calibration of Microwave Equipment. Calibrate microwave equipment In accordance with manufacturer's Instructions. If calibration Instructions are not available, see EPA Method 3051 [11]. 2. All digestion vessels and volumetric ware must be carefully acid washed and rinsed with reagent water. All digestion vessels should be cleaned by leaching with hot (1:1) nitric acid for a mItMmum of fifteen minutes, rinsed with reagent water, and dried In a clean environment. 3. Sample Digestion a. Tare the Teflon PFA digestion vessel. b. Weigh out 0.1 g paint chip sample to the nearest 0.001 g into the tared Teflon PFA sample ves­ sel. With large paint chip samples, measure out a 2 apiece, weigh to (he nearest 0.001 g, end quantitatively transfer It to ths vessel. c. Add S.Q ± 0.1 mL concentrated nitric acid to the sample vessel in a toms hood. If a vigorous reaction occurs, allow the reaction to stop before capping the vessel. Cap the vessel and torque (he cap to 12 ft-lb (16 N-m) according to the manufacturers directions. The sample vessel may be connected to an overflow vessel using Teflon PFA connecting tubas. Place the vessels in the microwave carrousel. Connect the overflow vessels to the center well of the unit. d. Place the vessels evenly distributed In the turntable of (he microwave unit using groups of two, six, or 12 sample vessels. Any vessels containing 5 mL of nitric acid tor reagent blank purposes are counted as sample vessels. When fewer than the recommended number of samples are to be digested, i.e., three samples plus one blank, the remaining vessels should be filled with 5 mL of nitric add to achieve the full complement of vessels. This provides an energy balance since the microwave power absorbed is proportional to the total mass In the cavity [14]. Irradiate each group of samples to achieve a temperature of 180*C in five minutes at a pressure of SO psi. Continue to irradiate to achieve a temperature of 16Q*C at 100 psi after 25 minutes. Continue

NIOSH Manual of AnsMtcat Mathmis (NWAM), Fourth t<S«on

09/20/01 THU 18:20 [TX/RX NO 8707]

CITY_21MC100_011904223

NYC-WTC_000281393

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000281393Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

NYC-WTC_000070845 · Box DEP Box 61

Chain of custody form for samples submitted to the Westchester County Department of Labs and Research in Valhalla, NY.

Similar indexed content

Compare copies and versions →