NYC Law Department release
Independent mirror · updated as the City releases more · v1.5.1
☕ Buy me a coffee
← Document results/Unboxed/No folder label
Document / 9 pages

ICP Elements by ICP ICP analysis, 7300

Machine-extracted title · confidence 22%

ICP element list and method for analysis; August 2004 issue reference.

NYC-WTC_000281368–000281376
Page image
Page image not rendered yet.
The pipeline renders page images incrementally; check back or read the OCR text alongside.
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document SEP.20.2001 3:27PM NO.432 P.9/14

ELEMENTS (ICPV, METHOD 7300, ISSUE 2, dated 15 AUOUSt 1994 • Page 8 of 9

APPENDIX > MICROWAVE DIGESTION POR LEAD IN PAINT CHIPS (AND OTHER MATRICES)

This procedure is an alternative to the procedure presented In the Sample Preparation section of this method, It provides a rapid, complete acid digestion prior to analysis by flame atomic absorption (PAA), heated graphite furnace atomic absorption (HGFAA), and inductively coupled plasma spectroscopy (ICP) 110]. A{2[2fQ|{usjQ^^Kjj|terig|{11-16] 1. Microwave apparatus requirements; a. The microwave unit provides programmable power with a minimum of 574 W and can be programmed to within ± 10 W of the required power. b. The microwave unit cavity is corrosion resistant as wall as ventilated. All electronics are protected against corrosion for safe operation. c. The system requires Teflon PFA digestion vessels (120-mL capacity) capable of withstanding pressures up to 7.5 ± 0.7 atm (110 ± 10 psi) and capable of controlled pressure relief at pressures exceeding 7.610.7 atm (110110 pd). d. A rotating turntable is employed to ensure homogeneous distribution of microwave radiation within ths unit. The speed of the turntable should be a minimum of 3 rpm. e. A safety concern relates to the use of saaled containers without pressure relief valves In ths unit Temperatura is ths Important variable controlling the reaction. Pressure Is needed to attain ele­ vated temperatures but must be safety contained [12]. f. Polymeric volumetric ware in plastic (Teflon or polyethylene), 50- or 100-mL capacity. g. Disposable polypropylene filter funnel. h. Analytical balance, 300-g capacity, and minimum ± 0.001 g.

Baaflaois 1, Nitric acid, oohcantratad, spectroscopy grade. 2. Reagent Water. Reagent water shall be interference frse. All references to water in the method refer to reagent water that meats the ASTM Type 2 standard.

Procedure

1. Calibration of Microwave Equipment. Calibrate microwave equipment In accordance with manufacturer's Instruetione. If calibration instructions are not available, see EPA Method 3051 [11]. 2. All digestion vessels and volumetric ware must be carefully add washed and rinsed with reagent water. All digestion vessels should be cleaned by leaching with hot (1:1) nitric acid for a minimum of fifteen minutes, rinsed with reagent water, and dried in a clean environment. 3. Sample Digestion a. Tare the Teflon PFA digestion vessel. b. Weigh out 0.1 g paint chip sample to the nearest 0,001 g into the tared Teflon PFA sample ves­ sel. With large paint chip samples, measure out a 2 cfhplece, weigh to the nearest 0.001 g, and quantitatively transfer It to the vessel. c. Add 5,0 ± 0.1 mL concentrated nitric acid to the sample vessel in a fUma hood. If a vigorous reaction occurs, allow the reaction to slop before capping the vessel. Cap the vessel and torque the cep to 12 ft-lb (16 N-m) according to the manufacturer's directions. The sample veesel may be connected to an overflow vessel using Teflon PFA connecting tubes. Place the vessels in the microwave carrousel. Connect the overflow vessels to the canter well of the unit. d. Place the vessels evenly distributsd In the turntable of the microwave unit using groups of two, six, or 12 sample vessels. Any vessels containing 5 mL of nitric add for reagent blank purposes are counted as sample vessels. When fewer than the recommended number of samples are to be digested, i.e., three samples plus one blank, the remaining vessels should ba filled with 5 mL of nitric acid to achieve the full complement of vessels. This provides an energy balance since the microwave power absorbed is prapohional to the total mass In the cavity [14], Irradiate each group of samples to achieve a temperature of 1S0'C in fve minutes at a pressure of 50 pal. Continue to irradiate to achieve a temperature of 1B0'C at 100 psi after 25 minutes. Continue

NIOSH Manual of AnaMloal Methods (NMAM), Fourth Edition

1 09/20/01 THU 18:20 [TX/RX NO 87071

CITY_21MC100_008110324

NYC-WTC_000281375

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000281375Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

ICP Elements, Aug 2004

NYC-WTC_000281386

Elements list for ICP analysis; includes aluminum, chromium, etc.

Similar indexed content

NYC-WTC_000070845 · Box DEP Box 61

Chain of custody form for samples submitted to the Westchester County Department of Labs and Research in Valhalla, NY.

Similar indexed content

Compare copies and versions →