Pace Analybcal Services, Inc
1700 Elm Street - Suite 200
Minneapolis, MN 55414
weAnaIvticar__________ Method 8290 Laboratory Control Spike Results
Tel- 612-607-1700
Fax: 612- 607-6444
Client - ATC ASSOCIATES
Lab Sample ID SPIKE-2232
Filename U21008B_01 Matrix WIPE
Total Amount Extracted 1.00 m2 Dilution NA
ICAL Date 08/06/2002 Extracted 09/29/2002
CCal Filename(s) U21008A_17 & U21008B_16 Analyzed 10/08/2002 21:33
Method Blank ID BLANK-2231 Injected By MRO
Native Qs Qm % Internal ng's Percent
Isomers (ng) (ng) Rec Standards Added Recovery
2,3,7.8-TCDF 0.20 0.19 97 2,3.7,8-TCDF-13C 2 00 76
2,3,7,8-TCDD-13C 2 00 75
1,2,3,7,8-PeCDF-13C 2 00 81
2.3,7.8-TCDD 0.20 0.19 96 2.3,4,7,8-PeCDF-13C 2.00 82
-1 1,2,3,7,8-PeCDD-13C 2 00 80
1,2,3,4.7.8-HxCDF-13C 2 00 86
1,2.3,7,8-PeCDF 1.00 1.00 100 1,2,3,6,7.8-HXCDF-13C 2.00 91
2,3,4,7,8-PeCDF 1 00 0 98 98 2,3,4,6.7,8-HxCDF-13C 2 00 87
1,2,3,7,8,9-HxCDF-13C
1,2,3,4,7,8-HxCDD-13C
2 00
2.00
87
94 I
1.2,3.7.8-PeCDD 1.00 0 99 99 1,2,3,6,7,8-HxCDD-13C 2 00 81 ♦
1,2,3,4,6,7,8-HpCDF-13C 2.00 77
1,2,3,4,7,8,9-HpCDF-13C 2.00 73
1,2,3,4,7.8-HxCDF 1.00 0.97 97 1,2,3,4,6,7.8-HpCDD-13C 2 00 69
1.2,3.6,7,8-HxCDF 1.00 0.84 84 OCDD-13C 4 00 63
2,3,4,6,7,8-HxCDF 1 00 1.00 100
1,2.3,7,8.9-HxCDF 1 00 1.00 100 1,2,3,4-TCDD-13C 2 00 NA
1.2,3,7.8,9-HxCDD-13C 2 00 NA
1.2,3,4,7,8-HxCDD 1.00 1 00 100 2,3,7,8-TCDD-37CI4 0.20 79
1,2,3,6,7,8-HxCDD 1.00 0.97 97
1,2,3,7,8,9-HxCDD 1 00 1.02 102
1,2,3.4,6,7,8-HpCDF 1.00 1.03 103
1,2,3,4,7,8,9-HpCDF 1.00 1.03 103
1,2,3,4.6,7,8-HpCDD 1.00 1 17 117
OCDF 2.00 2.52 126
OCDD 2.00 3.68 184 P
J
Qs = Quantity Spiked
Qm = Quantity Measured
Rec = Recovery (Expressed as Percent)
P = Recovery outside of target range
X - Background subtracted value
Nn = Value obtained from additional analysis
NA = Not Applicable
* = See Discussion Report No. ...1061946
REPORT OF LABORATORY ANALYSIS
This report shall not be reproduced, except in full,
without the wntten consent of Pace Analytical Services, Inc
CONFIDENTIAL, CITY_21MC100_0 03431276
SUBJECT TO PROTECTIVE ORDER NO. 2, SEPT. 2008
NYC-WTC_000239702
OCR can misread numbers and units. Confirm readings against the page image before using them.