Method 200.8 - Summary Report
Sample ID: QCS
Sample Date/Time: Wednesday, December 26, 2001 10:37:56
Sample Type: Sample
Sample Description: 50 PPB
Number of Replicates: 3
Batch ID:
Method File: d:\Elandata\Method\NEWE200.8.mth
Dataset File: d:\Elandata\Dataset\daily performance\QCS.1376
Sample Prep Volume (mL):
Initial Sample Quantity (mg):
Aliquot Volume (ml):
Diluted To Volume (mL):
Concentration Results
Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit
r Be 9 28720.583 0.248 0.040 50.06933 2.00 ug/L
|> Sc-1 45 715826.414 1.890 715826.414 ug/L
I V 51 763914.203 2.618 1.066 50.31003 3.55 mg/L
L Mn 55 1041294.357 0.724 1.455 50.61271 2.61 ug/L
r> In 115 1603839.564 0.246 1603839.564 ug/L
I Sb 121 522824.311 0.709 0.326 50.52043 0.81 ug/L
L Sb 123 402041.108 0.462 0.251 50.29357 0.24 ug/L
r> Ho 165 1915414.222 1.203 1915414.222 ug/L
L Tl 205 1706246.433 0.297 0.890 52.37856 1.09 ug/L
QC Calculated Values
Analyte Mass QC Std % Recovery Int Std % Recovery Spike % Recovery Dilution % Diff Dup. Rei. % Diff
r Be 9
|> Sc-1 45 97.745
I V 51
L Mn 55
r> In 115 96.319
I Sb 121
L Sb 123
r> Ho 165 100.928
L Tl 205
NYC-WTC_000085078
OCR can misread numbers and units. Confirm readings against the page image before using them.
NYC-WTC_000085078Source: NYC Law Department, mirrored locally
ICP-MS analysis of water samples using Method 200.8 with elemental measurements (Be, Sc, Mn, In, Sb, Ho) and quality control replicates performed on January 11, 2002.
1 shared topics · Same agency, volume and box · similarity 0.993
ICP-MS analysis of water samples using Method 200.8 with elemental measurements (Be, Sc, Mn, In, Sb, Ho) and quality control replicates performed on January 11, 2002.
ICP-MS analysis of water samples using Method 200.8 with elemental measurements (Be, Sc, Mn, In, Sb, Ho) and quality control replicates performed on January 11, 2002.