Daily Performance Report
Sample ID: DPC
Sample Date/Time: Friday, March 22,2002 14:51:32
Sample Description: DAILY PERFORMANCE CHECK
Method File: d:\elandata\Method\DAILY PERF TO SAVE REPORT.mth
Dataset File: d:\elandata\Dataset\Daily\DPC.018
Tuning File: d:\elandata\Tuning\default.tun
Optimization File: d:\elandata\Optimizeldefault.dac
Dual Detector Mode: Pulse
Acq. Dead Time(ns): 55
Current Dead Time (ns): 55
Summary
Analyte Mass Meas. Intens. Mean Net Intens. Mean Net Intens. SD Net Intens. RSD
Mg 24.0 54297.3 54297.269 v 192.863 0.4
Rh 102.9 329047.9 329047.926 1185.986 0.4
In 114.9 452053.2 452053.170 / 1833.586 0.4
Pb 208.0 260789.5 260789.490 946.602 0.4
> Ba 137.9 392670.5 392670.487 1391.141 0.4
Ba++ 69.0 5527.6 0.014 0.000 0.8
> Ce 139.9
CeO 155.9
485738.8
13781.3
485738.849
0.028 J 1485.325
0.001
0.3
1.8
Bkgd 220.0 8.8 8.833 1.034 11.7
Current Optimization File Data
Current Value Description
0.90 Nebulizer Gas Flow
6.00 Lens Voltage
1000.00 ICP RF Power
-1900.00 Analog Stage Voltage
1600.00 Pulse Stage Voltage
70.00 Discriminator Threshold
-6.00 AC Rod Offset
60.00 Service DAC 1
0.00 Quadrupole Rod Offset
Current Autolens Data
Analyte Mass Num of Pts DAC Value Maximum Intensity
Be 9 21 5.3 4267.0
Co 59 21 5.5 128777.0
In 115 21 6.5 408667.2
Sample ID: DPC
Report Date/Time: Friday, March 22,200214:54:58