NYC-WTC_000054548
Document page from the Department of Environmental Protection ICPMS series dated May 30, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 24, 2002.
NYC-WTC_000054375–000054546Folder label: “ICPMS 5/24/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document
WESTCHESTER COUNTY DEPT OF LABS AND RESEARCH Method 200.8 - Summary Report Sample ID: CCV STD2 60+5 PPB Sample Date/Time: Friday, May 24, 2002 13:35:05 Sample Type: Sample Sample Description: Number of Replicates: 3 Method File: c:\winnt\profiles\lrd1\elandata\Method\_NEWN.mth Dataset File: c:\winnt\profiles\Ird1\elandata\Dataset\new york city\CCV STD2 60+5 PPB.403 Aliquot Volume (mL): Diluted To Volume (mL):
Concentration Results Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit Be 9 23089.183 0.975 0.415 59.33046 0.23 ug/L Li 6 55653.400 1.098 55653.400 ug/L > Se-1 45 395145.456 1.289 395145.456 ug/L Cr 52 510840.104 0.200 1.277 61.60902 1.13 ug/L Cr 53 60942.250 2.066 0.153 61.99603 1.02 ug/L Mn 55 839118.436 2.487 2.122 61.71528 1.84 ug/L Co 59 595637.805 1.466 1.507 61.82433 1.52 ug/L Ni 60 121667.079 1.527 0.308 61.74017 1.06 ug/L Cu 63 256101.041 1.131 0.648 61.25207 0.43 ug/L Cu 65 121860.027 1.825 0.308 59.82596 1.08 ug/L Zn 66 83917.385 0.701 1.030 61.99892 0.27 ug/L Zn 67 14704.363 0.261 0.180 61.36344 0.53 ug/L Zn 68 62645.077 1.115 0.769 62.14385 0.58 ug/L > Ge 72 80842.004 0.589 80842.004 ug/L As 75 104296.823 0.775 1.138 55.88492 0.21 ug/L Se 77 10018.583 1.529 0.105 55.25458 1.72 ug/L Se 82 11576.153 1.540 0.143 62.79504 1.45 ug/L As-1 75 97576.854 0.563 97468.853 51.91364 0.56 ug/L Kr 83 719.036 3.288 -4028.252 ug/L Ag 107 502458.824 0.362 0.456 58.00215 1.32 ug/L Cd 111 136396.907 2.046 0.124 59.37061 0.37 ug/L Cd 114 320558.969 2.343 0.291 59.76177 2.25 ug/L > In 115 1100316.778 1.671 1100316.778 ug/L Sb 121 452800.753 1.053 0.412 61.13081 0.78 ug/L Sb 123 347026.156 0.631 0.315 61.07833 1.20 ug/L Ba 135 126459.192 1.458 0.078 59.95702 1.07 ug/L Ba 137 220854.119 0.720 0.137 59.91854 1.06 ug/L Ho 165 1636138.878 0.158 -0.012 ug/L > Tb 159 1613500.069 0.939 1613500.069 ug/L Hg 201 5849.273 1.775 0.004 5.03323 2.16 ug/L Hg 202 13446.409 2.603 0.008 5.03427 1.87 ug/L TI 205 1484328.406 1.275 0.920 59.38317 1.61 ug/L Pb 208 2029959.858 1.403 1.257 59.44293 0.68 ug/L
Sample ID: CCV STD2 60+5 PPB Report Date/Time: Friday, May 24, 2002 13:37:44 Page 1
NYC-WTC_000054544
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
NYC-WTC_000054548
Document page from the Department of Environmental Protection ICPMS series dated May 30, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054157
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 22, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053706
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 16, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054670
Document page from the Department of Environmental Protection ICPMS series dated May 31, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054747
Westchester County Department of Labs and Research report detailing elemental analysis results for sample AE12151 using ICPMS Method 200.8.
1 shared topics · Same agency, volume and box · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054886
Westchester County Department of Labs and Research report detailing elemental analysis results for sample ID [name] using ICPMS Method 200.8.
1 shared topics · Same agency, volume and box · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053630
Laboratory analysis results for a Westchester County Department of Labs and Research sample processed on May 15, 2002.
1 shared topics · Same agency, volume and box · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053505
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 15, 2002.
1 shared topics · Same agency, volume and box · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053212
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 9, 2002.
Same agency, volume and box · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
Same agency, volume and box · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077357
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar document embedding; no shared entity or topic indexed. · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
NYC-WTC_000054548
Document page from the Department of Environmental Protection ICPMS series dated May 30, 2002.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054157
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 22, 2002.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054157
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 22, 2002.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000054374
Document containing ICPMS analytical data or results from May 24, 2002.
Filed in the same folder
NYC-WTC_000054548
Document page from the Department of Environmental Protection ICPMS series dated May 30, 2002.
Similar indexed content
NYC-WTC_000054157
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 22, 2002.
Similar indexed content
NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
Similar indexed content
NYC-WTC_000053706
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 16, 2002.
Similar indexed content