NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
NYC-WTC_000040976–000041063Folder label: “ICPMS 4/23/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document
Daily Performance Report Sample ID: DPC Sample Date/Time: Tuesday, April 23,2002 08:56:59 Sample Description: DAILY PERFORMANCE CHECK Method File: d:\elandata\Method\DAILY PERF TO SAVE REPORT.mth Dataset File: d:\elandata\Dataset\Daily\DPC.052 Tuning File: d:\elandata\Tuning\default.tun Optimization File: d:\elandata\Optimize\default.dac Dual Detector Mode: Pulse Acq. Dead Time(ns): 55 Current Dead Time (ns): 55
Summary Analyte Mass Meas. Intens. Mean Net Intens. Mean Net Intens. SD Net Intens. RSD Mg 24.0 37557.4 37557.367 551.352 1.5 Rh 102.9 243079.3 243079.300 2314.230 1.0 In 114.9 338267.7 338267.726 2348.603 0.7 Pb 208.0 194573.1 194573.069 718.258 0.4 > Ba 137.9 315413.6 315413.603 2204.853 0.7 Ba++ 69.0 4620.2 0.015 0.000 1.0 > Ce 139.9 385293.6 385293.599 / 2620.241 0.7 CeO 155.9 11080.9 0.029 • 0.000 0.8 Bkgd 220.0 9.3 9.267 1.038 11.2
Current Optimization File Data Current Value Description 0.88 Nebulizer Gas Flow 6.00 Lens Voltage 1000.00 ICP RF Power -1900.00 Analog Stage Voltage 1600.00 Pulse Stage Voltage 70.00 Discriminator Threshold -6.00 AC Rod Offset 60.00 Service DAC 1 0.00 Quadrupole Rod Offset
Current Autolens Data Analyte Mass Num of Pts DAC Value Maximum Intensity Be 9 21 5.3 5796.8 Co 59 21 6.0 161848.0 In 115 21 6.5 475180.2
Sample ID: DPC Report Date/Time: Tuesday, April 23,2002 09:00:25 Page 1
NYC-WTC_000040976
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
No available related records indexed in this group.
NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077433
Summary of water sample results for metals analysis using Method 200.8 conducted on March 21, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077357
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077278
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077205
Summary of water sample results for metals analysis using Method 200.8 conducted on March 15, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000076792
Summary of water sample results for metals analysis using Method 200.8 conducted on March 6, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053212
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 9, 2002.
1 shared topics · similarity 0.990
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077560
Laboratory report from Westchester County Dept of Labs and Research detailing Method 200.8 metal concentrations in water samples dated March 21, 2002.
1 shared topics · similarity 0.987
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053630
Laboratory analysis results for a Westchester County Department of Labs and Research sample processed on May 15, 2002.
1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053505
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 15, 2002.
1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054375
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 24, 2002.
1 shared topics · similarity 0.985
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
1 shared topics · similarity 0.985
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
Page similarity is temporarily unavailable or this page has no indexed vector.
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000040975
Document containing ICPMS analytical data for the World Trade Center site from April 2002.
Filed in the same folder
NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 21, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar indexed content