NYC 9/11 Public Portal Document
Quantitation Report (QT Reviewed)
Data File C:\HPCHEM\l\DATA\MAR0102\2603.D Vial: 13
Acq On 1 Mar 2002 8:10 pm Operator:
Sample gc/ms scan 2/5 Inst : GC/MS 209
Mise Multiplr: 1.00
MS Integration Params: GOOFY.P
Quant Time: Mar 4 11:31 2002 Quant Results File: GCMS0208.RES
Quant Method : C:\HPCHEM\1\METHODS\GCMS0208.M (RTE Integrator)
Title : 209 625/TCLP calibration table
Last Update : Wed Feb 27 11:30:15 2002
Response via : Initial Calibration
DataAcq Meth : GCMS0208
Internal Standards R.T. Qlon Response Cone Units Dev(Min)
1) 1,4-Dichlorobenzene-d4 12.31 152 283527 20.00 ug/1 -0.04
10) Naphthalene-d8 15.96 136 1107908 20.00 ug/1 -0.04
17) Acenaphthene-diO 21.26 164 594601 20.00 ug/1 -0.05
29) Phenanthrene-diO 25.72 188 857428 20.00 ug/1 -0.05
38) Chrysene-dl2 33.79 240 602060 20.00 ug/1 -0.06
44) Perylene-dl2 38.06 264 415245 20.00 ug/1 -0.07
System Monitoring Compounds
37) 2,4-DDD 30.79 235 58574 6.54 ug/mL 0.30
Spiked Amount 5.000 Recovery = 130.80%
Target Compounds Qvalue
2) N-nitroso-dimethyl amine 0.00 74 0 N.D.
3) bis(2-Chloroethyl)ether 0.00 93 0 N.D.
4) 1,3-Dichlorobenzene 0.00 146 0 N.D.
5) 1,4-Dichlorobenzene 0.00 146 0 N.D.
6) 1,2-Dichlorobenzene 0.00 146 0 N.D.
7) 2,21 -oxybis(1-Chloropropan 0.00 45 0 N.D.
8) N-Nitroso-di-n-propylamine 0.00 70 0 N.D.
9) Hexachloroethane 0.00 117 0 N.D.
11) Nitrobenzene 0.00 77 0 N.D.
12) Isophorone 0.00 82 0 N.D.
13) bis(2-Chloroethoxy)methane 0.00 93 0 N.D.
14) 1,2,4-Trichlorobenzene 0.00 180 0 N.D.
15) Naphthalene 0.00 128 0 N.D.
16) Hexachlorobutadiene 0.00 225 0 N.D.
18) Hexachlorocyclopentadiene 0.00 237 0 N.D.
19) 2-Chloronaphthalene 0.00 162 0 N.D.
20) Dimethylphthalate 0.00 163 0 N.D.
21) 2,6-Dinitrotoluene 0.00 165 0 N.D.
22) Acenaphthylene 0.00 152 0 N.D.
23) Acenaphthene 0.00 153 0 N.D.
24) 2,4-Dinitrotoluene 0.00 165 0 N.D.
25) Diethylphthalate 22.74 149 528 N.D.
26) 4-Chlorophenyl-phenylether 0.00 204 0 N.D.
27) Fluorene 0.00 166 0 N.D.
28) Azobenzen/ 1,2-Diphenylhyd 0.00 77 0 N.D.
#) = qualifier out of range (m) = manual integration
603.D GCMS0208.M Mon Mar 04 11:32:13 2002 Page 1
NYC-WTC_000005335
OCR can misread numbers and units. Confirm readings against the page image before using them.