Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - Rev 1 - January 27, 2006
Method Blanks 1/prep batch Mercury must Reclean, Accuracy/Bias-
be < 15 ng per reanalyze Contamination
trap and/or qualify
data
Calibration 1/10 samples Mercury must Reclean, Accuracy/Bias-
Blanks be < 50 pg reanalyze Contamination
and/or qualify
data
Field Spikes 4 Percent Reanalyze Accuracy/Bias
spikes/background recoveries 80- and/or qualify and Precision
phase followed by 120%; % RSD
1 spike/every other < 30 in
week background
phase
Duplicate Every sample, RPDs must be Reanalyze Precision
Injections blank, QC <10 when and/or qualify
positive results
for both
injections are
> 5xQE
Triplicate Every 10 samples % RSD < 10 Reanalyze Precision
Injections when positive and/or qualify
results for all
injections are
> 5xQE
Standard 1/10 samples Percent Recalibrate, Accuracy/Bias
Additions recoveries 85- reanalyze
115% and/or qualify
data.
Breakthrough 1/location during Back half trap Qualify data Accuracy/Bias
Check background phase results must be and/or change
< 2% of front subsequent
half trap sampling
results or < 5 volumes.
ng per trap
Field/Trip 1/day Mercury must Reclean, Accuracy/Bias-
Blanks be < 15 ng per reanalyze Contamination
trap and/or qualify
data.