io Ambient Air Monitoring Program
Quality Assurance Project Plan
130 Liberty Street - NY, NY
Section No. 8
Revision No.: 0
Date: August 2005
co
I ■D
o Lower Manhattan Development Corporation Page 43 of 116 c
o CT
o o'
to ■D
00 o
o 3.
CD 03
TableS-S^Measur^ment^erformanceCriteriaTable-PCDDs/PCDFsbvHRGC/HRA^ o
QC Sample or Activity DQI o
Frequency Measurement Performance Criteria Corrective Action o
c
Field Duplicates l/week<"’ *RPD < 40 when positive results for both Assess laboratory precision, qualify Precision-Overall 3
samples are > 5x QL data, and/or resample data. CD
*No situation where one result is detected
at > 5x QL and other result is not detected
Extraction Standards Every sample, Percent recoveries 50-120% for tetra-hexa Reanalyze and/or qualify data. Accuracy/Bias
(pre-extraction) blank, QC congeners and 40-120% for hepta and octa
congeners
Field Spikes Every sample, Percent recoveries 70-130% Reanalyze and/or qualify data. Accuracy/Bias
(pre- sampling) blank, QC
Laboratory Control Sample 1/ batch Percent differences between LCS RRF and Determine cause of problem, Accuracy/Bias
(BCS3) (10 samples) ICAL RRF < 20 reanalyze, and/or qualify data.
2/analytical RPD between 2 LCS analyses < 20 Determine cause of problem, Precision
sequence reanalyze, and/or qualify data.
Method Blanks 1/ batch No target compounds > QL Reclean, retest, reanalyze and/or Accuracy/Bias-
qualify data. Contamination
Field/Trip Blanks l/day<’> No target compounds > QL Reclean, restest, reanalyze, and/or Accuracy/Bias -
qualify data. Contamination
O Pl IF Media Cleaning 1/batch cleaned No target compounds > QL Reclean and retest media. Accuracy/Bias-
Verification Check Contamination
Recovery Standards Every sample, Signal: noise ratio must be > 10:1 Reanalyze and/or qualify data. Accuracy/Bias
O blank, QC
m
Data Completeness Check NA Field 80%, Laboratory 95% NA Data Completeness
' 'InclLidcs the following compounds:
o ”C14-2,3,7,8-TCDD: 1.6ng
"Ci2-l,2,3,4,7-PeCDD; 4ng
J]
□0 ‘’Ci2-1,2,3,4,6.-PcCDF: 4ng
‘’Ci2-1,2,3,4,6,9-HxCDF: 4ng
‘’Ci2-l,2,3,4,6,8,9-HpCDF: 4ng
s
> ’^Yield duplicates may not be consistently collected at a frequency of 1/week during Phases 1 and 11 as the location selected for analysis will be dictated by PMio measurements.
^^’Will be collected at a frequency of 1/day; during Phase 1 and 11 only 1 field/trip blank per week will be analyzed.