PCB extracts for Aroclor analysis will be concentrated to a final volume of 5 mL. If the
total PCB analysis yields a result which exceeds the USEPA Site-Specific Trigger Level,
a further congener analysis will be performed on the same extract to confirm the total
PCB concentration. This analysis will also be performed using GC/ECD; however,
further QC analyses for the PCB congener analysis will not be required assuming the
results of the PCB Aroclor QC measurements (Table 8-4) demonstrate adequate accuracy
and precision of the sample extract analysis.
11.2 Analytical Quality Control
Tables 8-2 through 8-9 summarize the QC procedures checks, and samples, and their
respective acceptance limits for each fixed laboratory analytical parameter that will be
used during the project.
11.2.1 Field Analytical QC
Calibration procedures discussed in Section 13.2.1 will be adhered to for field analyses
of gaseous mercury in order to ensure the accuracy of sample measurements.
11.2.2 Fixed Laboratory QC
All required QC checks and QC samples and the associated QC acceptance limits are
detailed in the associated methods and in Tables 8-2 through 8-9.
11.2.2.1 Method Blanks/Preparation Blanks Method blanks will be performed as part of
each analytical batch for each methodology performed. Method blanks are used to
evaluate contamination introduced during sample preparation and/or analysis by the
laboratory.
11.2.2.2 Instrument Blanks Instrument blanks arc used to evaluate contamination
resulting from the analytical reagents and the instrumentation. In addition, instrument
blanks are sometimes used to assess potential carryover after the analysis of a highly
contaminated sample. Instrument blanks are only required for select analytical
parameters.
11.2.2.3 Surrogate Spikes Surrogate spikes are used to evaluate extraction efficiency or
analytical bias on a sample by sample basis for organic parameters. Surrogate spikes are
added to all samples for organic parameters. Surrogate spikes are another measure of
sample-specific QC.
11.2.2.4 Laboratory Control Samples Laboratory control samples (LCSs) and LCS
Duplicates are used to evaluate almost all parameters for the ability of the laboratory to
accurately and precisely identify and quantitate target compounds in a reference matrix
when spiked at the mid range of the calibration curve at a known concentration using a