Part II - Environmental Community Monitoring Plan
133-135 Greenwich & 21-23 Thames Streets
Airtek Project 05-0701 - Rev 2 September 16, 2005
differences, may be relatively poor. For this reason, collocated measurements are selected for use
in the precision calculations only when both measurements are above 60 g/m ’ for PM2.5.
The accuracy check for manual methods involves checking the flow rate of the sampler each
calendar quarter.
3.5 Quality Assurance for Gaseous Pollutants
These are some of the measures Airtek staff take to ensure the quality of data from the continuous
gas analysis systems. A "monitoring system" for a given pollutant at a given site usually consists
of an inlet tube (called a "probe"), an automatic electronic analyzer and a data acquisition system
(possibly connected to multiple instruments) and/or a strip chart recorder.
• Traceability of standards:
o Gaseous pollutant concentration standards (in compressed-gas cylinders or in
permeation tubes) traceable to a NlST-traceab1e reference material (NTRM) or to
a NIST-certified gas manufacturer’s internal standard (GMIS)
o Ultraviolet photometers (for ozone) certified against a NIST-certified reference
UV photometer at EPA
o mass flow meters (inside multigas dilution/gas phase titration calibrators)
certified traceable to NIST standards
o flow calibrators (bubble meters and/or dry flow calibrators) certified NIST-
traceable
o temperature, pressure, etc. certified NIST-traceable
• Multipoint calibrations: The sample collection staff notes the monitoring system's
response to multiple known concentrations, adjusting the system if necessary to read
correctly. This is done quarterly, after maintenance or repair and when a zero/span check
is out of range.
• Zero/span checks: The sample collection staff or an automated system records the
monitoring system's response to pure air, then to a concentration near its upper range
limit (at least once every 2 weeks).
• Precision checks: The sample collection staff or an automated system records the
monitoring system's response to a known concentration of a roughly consistent value
(every 2 weeks). (In practice, this is usually combined with the zero/span check.)
• Performance audits:
o Quarterly audits: The IH Quality coordinator using a different calibrator and
different gas standards (or a different photometer) notes the monitoring system's
response to multiple concentrations. Normally, this is done quarterly for each
monitor.
• Data audit: Airtek Data Manager (not the operator of the monitoring system in
question) verifies the highest and one random hourly reading each day for each
parameter, and reviews all the data for anomalies.