QAPP AMENDMENT FORM: Issue 2 DATE FORM SUBMITTED: 4/5/06
QAPP Title Quality Assurance Project Plan for the Ambient Air Monitoring Program 130 Liberty Street Deconstruction Project September 7, 2005
How rate checks and calibrations associated with the high volume samplers used for metals will be revised as follows:
• A magnehelic gauge will be permanently installed to the side of the sampler for calibrations and daily flow rate checks. This will replace the orifice previously used. • The requuement for the weekly flow rate check will be removed. Currently, the samplers are calibrated once per quarter and/or after maintenance. The flow rates are checked before and after sampling on a daily basis via the magnehelic readings. • A criterion of < 20 will be introduced for the relative percent difference (RPD) between the initial and final magnehelic readings. • Calculation of flow rates on a daily basis will be revised to correct for the average temperature and pressure readings during the sampling period. Reason for Amendment: Since the flow rates are verified on a daily basis, the weekly flow checks are redundant and therefore have been eliminated. In order to assess the precision of sample collection, the RPD criterion for initial and final magnehelic readings has been introduced as a means of evaluating initial and final flow differences. This evaluation will serve as an additional measure to verify the accuracy of the reported volumes. If RPDs are greater than 20, associated data will be flagged as estimated. The flow rate calculations will be corrected for the average temperature and pressure readings during the sampling period so as to be consistent with the flow rate calculations currently in place for all other parameters collected as part of this program. Sections of QAPP Affected: 10.2.1, Metals (TSP Filters): A magnehelic gauge will be added to the list of required equipment. 13.1.1, Field Equipment, Table 13-la: The weekly flow check of the high volume samplers will be eliminated and replaced with daily checks of the flow rate via the magnehelic readings. An RPD criterion of < 20 will be introduced as a means of evaluating the initial and final magnehelic readings. The corrective action will be to flag the associated data as estimated if the RPD criterion is exceeded. 13.2.1, Field Equipment, Table 13-3a: The weekly flow check of the high volume samplers will be eliminated and replaced with daily checks of the flow rate via the magnehelic readings. An RPD criterion of < 20 will be introduced as a means of evaluating the initial and final magnehelic readings. The corrective action will be to flag the associated data as estimated if the RPD criterion is exceeded. Attachment C, SOP for TE-5170V Total Suspended Particulate VFC High Volume Air Sampler for Metal^ The SOP has been revised to reflect the edits discussed above. In addition, the calibration frequency has been revised to reflect the frequencies cited in the QAPP. Details of the calibration procedure, including all required equations, and calculation of sample flow rates have been added to the SOP. Calculation of sample flow rates has been revised to include correction based on the average temperature and pressure reading during the sampling period. This SOP has been reissued as Revision 1, February 2006. Date Implemented: January 2006 (Phase I)______________________________________________________________________
Page 1 of 7
CITY_DEP_WORBY_CANNATA_SUBP_000078982
NYC-WTC_000425341
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
NYC-WTC_000451394
QAPP amendment detailing flow rate and calibration changes for high-volume samplers, submitted 2006.
16 shared non-person entities · 1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000427863
QAPP amendment to raise mercury spike from 50 ng to 150 ng for improved recoveries.
6 shared non-person entities · 1 shared topics · similarity 0.923
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000425322
QAPP amendment modifying mercury spike level to 150 ng for field samples, Oct 2005.
6 shared non-person entities · 1 shared topics · similarity 0.922
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000446799
QAPP amendment documenting mercury spike level increase from 50 to 150 ng.
6 shared non-person entities · 1 shared topics · similarity 0.920
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000441952
QAPP amendment changing mercury spike level from 50 to 150 ng in 2005.
6 shared non-person entities · 1 shared topics · similarity 0.920
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000423693
QAPP amendment increasing mercury spike level from 50 to 150 ng/trap for improved accuracy.
7 shared non-person entities · similarity 0.919
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000443742
Chain-of-custody form for Environmental Community Air Monitoring Program plans at Greenwich Street and Thames Street, dated November 4, 2005.
17 shared non-person entities · similarity 0.918
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000444945
Plan and submittals for environmental community air monitoring at 133-135 Greenwich and 21-23 Thames.
17 shared non-person entities · similarity 0.918
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000427582
CoC document for Greenwich/Thames Street air monitoring plan dated Nov 9, 2005.
17 shared non-person entities · similarity 0.918
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000440780
Plan document for environmental air monitoring at Greenwich Street and Thames Street sites, dated November 4, 2005.
17 shared non-person entities · similarity 0.918
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000426348
Plan document outlining community air monitoring for Greenwich and Thames sites.
17 shared non-person entities · similarity 0.918
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000427716
CoC document for Greenwich/Thames Street air monitoring plan dated Nov 9, 2005.
17 shared non-person entities · similarity 0.918
machine-extracted · confidence unavailable · Check page ↗No available related records indexed in this group.
Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
NYC-WTC_000451394
QAPP amendment detailing flow rate and calibration changes for high-volume samplers, submitted 2006.
Similarity score 0.995 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000451394
QAPP amendment detailing flow rate and calibration changes for high-volume samplers, submitted 2006.
Similarity score 0.960 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000421711
Quality assurance plan for ambient air monitoring during LMDC project at 130 Liberty Street.
Similarity score 0.948 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000426821
QA plan reference for deconstruction ambient air monitoring, Sept 2005.
Similarity score 0.948 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000421972
QA plan for ambient air monitoring for the 130 Liberty Street deconstruction project, Sept 2005.
Similarity score 0.948 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000423219
Quality assurance plan and project details for ambient air monitoring near 130 Liberty Street, Sep 2005.
Similarity score 0.948 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Machine-extracted building matches; verify each source. Buildings, never households.
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000451394
QAPP amendment detailing flow rate and calibration changes for high-volume samplers, submitted 2006.
Similar indexed content
QAPP amendment to raise mercury spike from 50 ng to 150 ng for improved recoveries.
Similar indexed content
NYC-WTC_000425322
QAPP amendment modifying mercury spike level to 150 ng for field samples, Oct 2005.
Similar indexed content
QAPP amendment documenting mercury spike level increase from 50 to 150 ng.
Similar indexed content
QAPP amendment changing mercury spike level from 50 to 150 ng in 2005.
Similar indexed content