Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
verification of the field analysis records by the AIRTEK Field Sampling Coordinator,
or designee, will ensure that:
• Logbooks and standardized forms have been filled out completely and that the
information recorded accurately reflects the activities that were performed.
• Records are legible and in accordance with good record keeping procedures,
i.c., entries arc signed and dated, data arc not obliterated, changes arc initialed,
dated, and explained.
• Calibration procedures were conducted in accordance with the protocols
described in the QAPP, and that any deviations were documented and
approved by the appropriate personnel.
16.1.3 Fixed Laboratory Data
16.1.3.1 Internal Reviews
Prior to the release of any data from the laboratory, the data will be verified and
approved by laboratory personnel. This review will consist of a tiered review by the
person performing the work, a qualified peer, and by supervisory personnel. Each
laboratory used in the program has a procedure in place for documenting all levels of
data review. Prior to being released as final, laboratory data will proceed through a
tiered review process. Data verification starts with the analyst or technician who
performs a 100 percent review of the data to ensure the work was done correctly the
first time. It is the responsibility of the analyst or technician to ensure that the
verification of data in his or her area is complete. The data reduction and initial
verification process must ensure that:
• Sample preparation and analysis information is correct and complete,
• Results arc correct and complete,
• The appropriate methods have been followed and are identified in the project
records,
• Proper documentation procedures have been followed,
• All nonconformances have been documented,
• Project-specific requirements have been met.
Following the completion of the initial verification by the analyst or technician, a
systematic check of the data will be performed by an experienced peer, Laboratory
Section Leader, or designee. This check will be performed to ensure that initial
review has been completed correctly and thoroughly. Included in this review will be
an assessment of the acceptability of the data with respect to:
• Adherence of the procedure used to the referenced methods and specific
instructions,
• Correct interpretation of data (e.g., mass spectra, chromatographic
interferences, etc.),