NYC 9/11 Public Portal Document

jnt by: IDAHO TECHNOLOGY 8015880507; 10/24/01 15:59; letFaji #583:Page 2/7

R.A.P.I.D. Instrument and Anthrax Reagent Validation Specificity and Sensitivity of the Anthrax Lethal Factor Assay

All assays arc independently tested at the Air Force Institute of Pathology (AFlP) for reproducibility of crossing points, precision, limit of detection, and analytical specificity against approximately 80 organisms. Assays are tested for clinical sensitivity and specificity with approximately 250 blood cultures and 250 clinical isolates from Walter-Reed Army Medical Center laboratories.

The R.A.P.I.D. B. anlhracis Lethal Factor Detection Kit is specially adapted for PCR in glass capillaries using the R.A.P.I.D. instrument and hybridization probes for sequence specific detection of B. anthracis Lethal Factor.

A fragment of B. anthracis Lethal Factor is amplified using specific primers. The amplicon is detected by fluorescence using a specific pair of Hybridization Probes. The Hybridization Probes consist of two different short oligonucleotides that hybridize to an internal sequence of the amplified fi’agment during the annealing phase of the PCR cycle. One probe is labeled at the 5’-end with LightCycler-Red 640, and to avoid extension, modified at the 3‘-end by phosphorylation. The second probe is labeled al the 3’-end with fluorescein. Only after hybridization to the template DNA do the two probes come in close proximity, resulting in fluorescence resonance energy transfer between the two fluorophores. The fluorescence emitted by the LightCycler-Red 640 dye is measured in the R.A.P.I.D. instrument in channel 2.

The fluorescent signal from the unknown sample is compared to the signals from the positive and negative control samples.

Specificity The R.A.P.I.D. - B. anthracis Lethal Factor Detection Kit has demonstrated specific detection of B. anlhracis Lethal Factor versus a panel of 84 pathogenic agents, both related and unrelated (see appendix C). Specificity was 100%. Only BA strains containing the protective antigen gene reacted with the reagent.

This assay recognizes the following strains ofE. anthracis

B. a. de|ta-Ames B. a. Vollum B. a. Zimbaburts B.a. GT 41 B.a. GT 34 B.a. GT B5 B.a. GT 29 B.a. GT 55 B.a. GT 30 B a. GT 52 B.a. GT 77 B.8.GT 80 B.a. GT 23 B.a. GT 38 B.a. GT 69

CITY_21MC100_002971229

NYC-WTC_000354382

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000354382Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

WTC health criteria appendix, Dec 2001 · Box unavailable

NYC-WTC_000176140

Appendix document outlining health protective screening values and contaminants of concern for WTC site workers and Lower Manhattan residents.

Similar document embedding; no shared entity or topic indexed. · similarity 0.865

machine-extracted · confidence unavailable · Check page ↗

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →