Pace Analytical Services, Inc.
1700 Elm Street, Suite 200
*aceAnalyticar Minneapolis, MN 55414
Phone: 612.607.1700
www.pacelabs.com REPORT OF: CHEMICAL ANALYSES Fax: 612.607.6444
' PROJECT: PCDD/PCDF ANALY SES DATE: January 9, 2002
PAGE: 5 REPORT NO: 01-1052540
- PCDD/PCDF Quantification and Calculations (Cont.)
Each pair of ion mass peaks in the selected-ion-current chromatograms was evaluated manually to
determine if it met the criteria for a PCDD or PCDF isomer. Areas of all peaks exhibiting correct ion
ratios, having retention times within the correct windows, and having areas corresponding to
concentrations in the range covered by the initial calibration were then summed for calculations of
total congener concentrations. The toxic equivalence of each sample was calculated using the
factors listed in Table 5. ,
■STS
A limit of detection (LOD) based on producing a signal that is 2.5 tirnes the noise level, was
calculated for each undetected 2,3,7,8-substituted isomer of any tetra through octa chlorinated
congener class. The noise heights used to calculate the detection limits were measured at the
retention time of the specific isomer. The formula used for calculating the LOD is:
-•
f“5
Hn X Qis X 2.5
ihl LOD =
His x,W X Rf
where;
- LOD Single isomer limit of detection
. Hn Sum of noise heights at native isomer retention time
: j Qis Quantity of labeled internal standard
His Sum of peak heights for labeled internal standard
W Sample amount
Rf Response factor
a
, The recovery of the 2,3,7,S-TCDD-^’Cl^ enrichment efficiency standard and each ^^CT2-labeled
j internal standard, relative to either 1,2,3,4-TCDD-’®Ci2 or 1,2,3,7,8,9-HxCDD-’^Ci2, was calculated
' using the following equation:
i Ais X Qrs x 1 00%
%R =
Rfr X Ars x Qis
- where:
, , %R = Percent recovery of labeled internal standard
I Ais = Sum of integrated areas of labeled internal standard L
Qrs = Quantity of recovery standard
, Ars = Sum of integrated areas of recovery standard
JI Rfr = Response factor of the specific labeled internal standard relative to the recovery standard
Qis = Quantity of the labeled internal standard congener added to the sample
REPORT OF LABORATORY ANALYSIS
■ “ This report shall not be reproduced, except in full,
without the written consent of Pace Analytical Services, Inc.