3 PCDD/PCDF Quantification and Calculations
1 i The PCDD/PCDF isomers were quantified by comparison of their responses to the responses of the
labeled internal standards. Relative response factors were calculated from analyses of standard
; I mixtures containing representatives of each of the PCDD/PCDF congener classes at five
41 concentration levels, and each of the internal standards at one concentration level, as shown in
Table 4. The PCDD/PCDF response factors were calculated by comparing the sum of the
P responses from the two ion masses monitored for each chlorine congener class to the sum of the
’ ’ responses from the two ion masses of the corresponding isotopically labeled internal standard. The
formula for the response factor calculation is:
An X Qis
Rf =
4. Ais X Qn
where;
Rf = Response factor
An = Sum of integrated areas for native isomer
Qis = Quantity of labeled internal standard
- Ais = Sum of integrated areas for labeled internal standard
Qn = Quantity of native isomer
■3
The levels of PCDD/PCDF in each sample were quantified using the following equation:
An X Qis.
C =
Ais x W X Rf
■ where:
, , C = Concentration of target isomer or congener class
_■ An = Sum of integrated areas for the target isomer or congener class
Qis - Quantity of labeled internal standard added to the sample
Ais = Sum of integrated areas for the labeled internal standard
} W = Sample amount
Rf - Response factor