PROJECT: PCDD/PCDF ANALYSES DATE: January 9, 2002
PAGE: 2 REPORT NO: 01-1052540
'Hi ip
PCDD/PCDF Analyte Enrichment
-
t The extraction procedure often removes a variety of compounds, in addition to the PCDDs and
PCDFs, from the sample matrix. Some of these compounds can directly interfere with the analyses
while others can overload the capillary column causing degradation in chromatographic resolution
Hl or sensitivity. The analyte enrichment steps described below are used to remove interferences
fromYhe extracts.
Each extract was diluted to 100 mL with hexane, transferred to a separatory funnel, and washed
with 1N sodium hydroxide, concentrated sulfuric acid, and aqueous sodium chloride (5% w/v) as
-“21
needed. The hexane extract was quantitatively transferred to a liquid chromatography column
* containing alternating layers of silica gel, 40% concentrated sulfuric acid on silica gel, and 33% 1
, N sodium hydroxide on silica gel. The column was eluted with 90 mL of hexane and the entire
I eluate was collected and concentrated, under ambient conditions, to a volume of 1 mL.
: ; Each extract was then fractionated on a liquid chromatography column containing 4 g of activated
/ ( alumina. The column was eluted with 20 mL of hexane followed by 15 mL of 60% methylene
chloride/hexane. The 60% methylene chloride/hexane fraction was concentrated to 1 mL under a
• J stream of dry nitrogen and applied to the top of a chromatography column containing 1 g of 5%
i I AX-21 activated carbon in silica gel. The column was eluted with two 2-mL portions of hexane, 2
mL of cyclohexane/methylene chloride (50:50 v/v) and cyclohexane/methanol/toluene (75:20:5
oj v/v) in the forward direction, and then with toluene in the reverse direction. The toluene fraction
(4 was collected, concentrated, spiked with recovery standards (1 ,2,3,4-TCDD-^^Ct2 and 1,2,3,7,8,9-
HxCDD-''®Ci2) and taken to a final volume of 20 uL.
PCDD/PCDF Analyses
■fl ■ ■
Lj Each sample extract was analyzed for the presence of PCDDs and PCDFs using combined
capillary column gas chromatography/high resolution mass spectrometry (HRGC/HRMS). The
; instrumentation consisted of a Hewlett Packard Model 6890 gas chromatograph interfaced to a
-3 Micromass Ultima high-resolution mass spectrometer. The capillary column was interfaced
directly into the ion source of the mass spectrometer, thus providing the highest possible
j sensitivity while minimizing degradation of the chromatographic resolution.