Attention: Mr Goran Berndtsson
Re: Calibration History of the Data lossins SE400 Kits
Dear Goran,
The Data Logging SE400 Kits that you currently have, are serial numbers:
Fan Serial Number Work Order % Error at % Error at
50x2 litres 25x2 litres
FO12295 2002-147-01-MF +7% +15%
FO12283 2002-147-01-MF +1% +13%
F012288 2002-147-01-MF +5% +17%
F012287 2002-147-01-MF +3% +9%
FO12293 2002-189-01-MF +3% +8%
FO12296 2002-189-01-MF +2% +5%
Originally when Data Logging Kits were developed, the components were all standard, approved components, with the
exception of the fan unit. The fan units were fitted with what were claimed to be improved transducers. It was hoped that
these new transducers would give more precise results than the standard NIOSH approved transducers.
After evaluating the performance of the new transducers it was determined that the performance was no better than the
standard NIOSH approved configuration.
A decision was made to rebuild all Data Logging Kits to NIOSH standard configuration.
The units listed above were rebuilt, to NIOSH standard configuration, and calibrated on the Work Orders shown above. The
flow calibration results achieved for each unit are also shown above for your information. The calibration was done at
between 20 and 25 degrees centigrade.
I hope that this letter addresses your requirements.
Sincerely,
Andrew Smith
Quality Assurance Manager
[email protected]
The SEA Group