Pace Analytical Services, Inc.
1700 E/m Street, Suite 200
'aceAnalytkaT Minneapolis, MN 55414
Phone: 612.607.1700
www.pacelabs.com REPORT OF: CHEMICAL ANALYSES Fax: 612.607.6444
PROJECT: PCDD/PCDF ANALYSES DATE: October 17, 2002
PAGE: 2 REPORT NO: 02-1061865
DISCUSSION (Cont.)
It should be noted that two of the samples (DX3 and DX4) were extracted beyond the 30 day
recommended hold time described in the method. Per method direction, the results should be
considered minimum levels. It should be emphasized that the hold times are recognized as
uncertain by the method and are considered recommendations. (Per Section 6.4 of Method 8290,.
hold times of up to 1 year may apply.) This is similar to Method 1613, a comparable dioxin
method, which recognizes a one year holding time. Based on laboratory experience, it is extremely
unlikely that the results would vary due to the additional time taken to complete the extraction of
the sample material.
The response factor for the labeled OCDD internal standard in two end of shift calibration
standards were outside the target range for this method. The response factors from the beginning
and end calibrations were averaged to provide an accurate value for this isomer, as described in the
method. The affected points are flagged on the data summary sheets.
Laboratory spike samples were also prepared with the sample batch by extracting clean wipes
that had been fortified with native standard materials. The results, found at the end of Appendix
B, show that the spiked native compounds were generally recovered at 84-126% with relative
percent differences of 0.0-21,1%. It should be noted that one native isomer was recovered
above the target range for this method and is flagged "P" on the data summary sheet. The
related RPD is also elevated for this isomer.
REMARKS
The sample extracts will be retained for a period of 30 days from the date of this report and then
discarded unless other arrangements are made. The raw mass spectral data will be archived on
magnetic tape for a period of not less than one year. Questions regarding the data contained in
this report may be directed to the authors at the numbers provided below.
Pace Analytical Services, Inc.
Unze XI
pi^ct Manager, Dilins
(612)607-6383
REPORT OF LABORATORY ANALYSIS
This report shall not be reproduced, except in full,
without the written consent of Pace Analytical Services, Inc.
mebci
CONFIDENTIAL, CITY_21MC100_003430427
SUBJECT TO PROTECTIVE ORDER NO. 2, SEPT. 2008
NYC-WTC_000238764