PCDD/PCDF Quantification and Calculations (Cont.)
Each pair of ion mass peaks in the selected-ion-current chromatograms was evaluated manually to
determine if it met the criteria for a PCDD or PCDF isomer. Areas of all peaks exhibiting correct ion
ratios, haying retention times within the correct windows, and having areas corresponding to
concentrations in the range covered by the initial calibration were then summed for calculations of
total congener concentrations. The toxic equivalence of each sample was calculated using the
factors listed in Table 5.
A limit of detection (LOD) based on producing a signal that is 2.5 times the noise level, was
calculated for each undetected 2,3,7,8-substituted isomer of any tetra through octa chlorinated
congener class. The noise heights used to calculate the detection limits were measured at the
retention time of the specific isomer. The formula used for calculating the LOD is:
Hn X Qis X 2.5
LOD =
His X W X Rf
where:
' LOD = Single isomer limit of detection
.. Hn = Sum of noise heights at native isomer retention time
Qis = Quantity of labeled internal standard
His = Sum of peak heights for labeled internal standard
W = Sample amount
Rf = Response factor
The recovery of the 2,3,7,S-TCDD-^CU enrichment efficiency standard and each ^’’Gi2-labeled
internal standard, relative to either 1,2,3,4-TCDO-i^Ci2 or 1,2,3,7,8,9-HxCDD-’5Ci2, was calculated
using the following equation:
Ais X Qrs x 100%
%R =
Rfr X Ars x Qis
where:
%R - Percent recovery of labeled internal standard
Ais Sum of integrated areas of labeled internal standard
Qrs Quantity of recovery standard
Ars Sum of integrated areas of recovery standard
Rfr Response factor of the specific labeled internal standard relative to the recovery standard
Qis Quantity of the labeled internal standard congener added to the sample
REPORT OF LABORATORY ANALYSIS
This report shall not be reproduced, except in full,
without the written consent of Pace Analytical Services, Inc.