Co-located samples will be collected in 10% of the air samples. While not a true duplicate,
evaluation of the results will provide some information on the quality of the sampling technique.
Generally, analysis of duplicate samples should be within +/- 25% of each other.
At least two field blanks (or 10% of samples submitted, whichever is greater) per air sample
method per sample team will be submitted with the air samples collected each day. Blank
cassettes should be opened at the same time as the sample cassettes are opened (just prior to
sampling). Blank cassettes will be recapped and stored in a clean area (e.g., closed bag or box)
with the covers from the sampling cassettes during the sampling period.
Each air sampling pump will be calibrated with a representative sampling device in-line before
and after sample collection. Sample flow rates will be documented in the sampling log. The
average of pre- and post-calibration flow rates will be used as the sample flow rate, for sample
volume calculations.
Dust samples will be collected using the U.S. Environmental Protection Agency (EPA)
Environmental Response Team (ERT) indoor dust sampling standard operating procedure (SOP).
Duplicate samples are not feasible with indoor surface dust sample collection. Co-located
samples (samples collected from adjacent areas of the floor) will be collected in 10% of the
indoor surface dust sample locations. While not a true duplicate, evaluation of the results will
provide some information on the variability of contaminants in the dust.
Samples should not be packaged in untreated polystyrene-electrostatic forces may cause fiber
loss from some sample filters.
Chain of custody will be established and maintained throughout sample collection, transport, and
analysis of the samples. Laboratory analysis shall follow all method-required quality assurance
and quality control procedures.
All samples collected, unless analyzed with destructive methods, will be archived until further
notice.