ASBESTOS and OTHER FIBERS by PCM: METHOD 7400, Issue 2. dated 15 August 1994 - Page 7 of 15
(1) Count as 1/2 fiber any fiber with only one end lying within the graticule area, provided
that the fiber meets the criteria of rule a above.
(2) Do not count any fiber which crosses the graticule boundary more than once.
(3) Reject and do not count all other fibers.
c. Count bundles of fibers as one fiber unless individual fibers can be identified by observing
both ends of a fiber.
d. Count enough graticule fields to yield 100 fibers. Count a minimum of 20 fields. Stop at
100 graticule fields regardless of count.
19. Start counting from the tip of the filter wedge and progress along a radial line to the outer edge.
Shift up or down on the filter, and continue in the reverse direction. Select graticule fields
randomly by looking away from the eyepiece briefly while advancing the mechanical stage.
Ensure that, as a minimum, each analysis covers one radial line from the filter center to the
outer edge of the filter. When an agglomerate or bubble covers ca. 1/6 or more of the graticule
field, reject the graticule field and select another. Do not report rejected graticule fields in the
total number counted.
NOTE 1: When counting a graticule field, continuously scan a range of focal planes by
moving the fine focus knob to detect very fine fibers which have become embedded
in the filter. The small-diameter fibers will be very faint but are an important
contribution to the total count. A minimum counting time of 15 seconds per field is
appropriate for accurate counting.
NOTE 2: This method does not allow for differentiation of fibers based on morphology.
Although some experienced counters are capable of selectively counting only fibers
which appear to be asbestiform, there is presently no accepted method for ensuring
uniformity of judgment between laboratories. It is, therefore, incumbent upon all
laboratories using this method to report total fiber counts. If serious contamination
from non-asbestos fibers occurs in samples, other techniques such as transmission
electron microscopy must be used to identify the asbestos fiber fraction present in
the sample (see NIOSH Method 7402). In some cases (i.e., for fibers with
diameters >1 pm), polarized light microscopy (as in NIOSH Method 7403) may be
used to identity and eliminate interfering non-crystalline fibers [15].
NOTE 3: Do not count at edges where filter was cut. Move in at least 1 mm from the edge.
NOTE 4: Under certain conditions, electrostatic charge may affect the sampling of fibers.
These electrostatic effects are most likely to occur when the relative humidity is low
(below 20%), and when sampling is performed near the source of aerosol. The
result is that deposition of fibers on the filter is reduced, especially near the edge of
the filter. If such a pattern is noted during fiber counting, choose fields as close to
the center of the filter as possible [5].
NOTE 5: Counts are to be recorded on a data sheet that provides, as a minimum, spaces on
which to record the counts for each field, filter identification number, analyst's name,
date, total fibers counted, total fields counted, average count, fiber density, and
commentary. Average count is calculated by dividing the total fiber count by the
number of fields observed. Fiber density (fibers/mm is defined as the average
count (fibers/field) divided by the field (graticule) area (mm ^/field).
CALCULATIONS AND REPORTING OF RESULTS
20. Calculate and report fiber density on the filter, E (fibers/mm ^), by dividing the average fiber count
per graticule field, F/n,, minus the mean field blank count per graticule field, B/n by the
graticule field area, A, (approx. 0.00785 mm
E = —------- fibers/mm 2.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94