Method 200.8 ICP-MS lab results, NEWE samples, Jan 11, 2002
Machine-extracted title · confidence 85%
ICP-MS analysis of water samples using Method 200.8 with elemental measurements (Be, Sc, Mn, In, Sb, Ho) and quality control replicates performed on January 11, 2002.
Method 200.8 - Summary Report
Sample ID: AD31674
Sample Date/Time: Friday, January 11, 2002 14:39:21
Sample Type: Sample
Sample Description:
Number of Replicates: 3
Batch ID:
Method File: d:\elandata\Method\NEWE200.8.mth
Dataset File: d:\elandata\Dataset\daily performance\AD31674.2051
Sample Prep Volume (mL):
Initial Sample Quantity (mg):
Aliquot Volume (mL):
Diluted To Volume (mL):
Concentration Results
Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit
r Be 9 10.333 65.873 -0.000 -0.00099 1482.70 ug/L
Sc-1 45 611137.154 2.439 611137.154 ug/L
L Mn 55 629329.997 1.868 1.029 35.67634 2.55 ug/L
n In 115 1284473.376 1.497 1284473.376 ug/L
I Sb 121 419.014 4.932 0.000 0.04645 3.95 ug/L
L Sb 123 335.483 2.698 0.000 0.04877 3.99 ug/L
n Tl Ho 165 1721809.156 1.286 1721809.156 ug/L
L 205 233.339 7.671 0.000 0.00096 70.07 ug/L
QC Calculated Values
Analyte Mass QC Std % Recovery I nt Std % Recovery Spike % Recovery Dilution % Diff Dup. Rei. % Diff
r Be 9
|> Sc-1 45 96.251
L Mn 55
r> In 115 97.399
I Sb 121
L Sb 123
r> Ho 165 95.867
L Tl 205
Calibration
Analyte MassCurve Type Correlation Coefficient
r Be 9Linear Thru Zero 0.999992
|> Sc-1 45Linear Thru Zero
L Mn 55Linear Thru Zero 0.999983
r> In 115Linear Thru Zero
I Sb 121 Linear Thru Zero 0.999998
L Sb 123Linear Thru Zero 1.000000
r> Ho 165Linear Thru Zero
L Tl 205Linear Thru Zero 0.999977