Sample 10187
The result for bis(2-ethylhexyl)phthalate is being reported as <1 ug/L.
The result generated by the data system (1.5 ug/L) was corrected because when the 1
ug/L calibration standard of 5/1/02 was quantitated using the calibration of 5/1/02 it
yielded a result for phthalate of 1.6 ug/L. This result varies from the true value of 1 ug/L
by more than 30%. Therefore, the phthalate result for this sample was approximated
using a one point calibration.
analysis phthalate area int std area computer generated
cone (ug/L)
1 ug/L cal std 214357 2773851 1.6
(5/1/02)
sample 10187 195093 3035920 1.5
RF = (214357)(5) = 0.3864
(2773851)(1)
C = (195093)(5) 0.83 ug/L, bis(2-ethylhexyl)phthalate (or <1 ug/L.)
(3035920)(0.3864)
Also phthalate is in the lab blank.
Benzo(a)pyrene (BAP) appears to be present in sample 10187 at approximately 0.29 ug/L
and, therefore, the result is being reported as such. Note: there is some background
interference present.
The result of 0.80 ug/L as generated by the data system was corrected as noted above.
The correction was made because when the 0.2 ug/L calibration standard of 5/1/02 was
quantitated using the calibration of 5/1/02 it yielded a result of of 0.79 ug/L for BAP.
This result varied by more than 30% from its true value.
Therefore, the BAP sample result was approximated based on a one point calibration
using the 0.2 ug/L calibration standard.
analysis BAP area internal standard area RF
(chrysene d-12)
0.2 ug/L 18392 3658766 0.1257
cal std of 5/1/02
sample 10187 22029 3035920
NYC-WTC_000082583