The recoveries for 2 of the 18 compounds in the LCS Standard were below their
acceptance ranges. The recoveries for 3 of the 18 compounds in the Spiked
Sample were outside of their acceptance ranges. The recovery for 1 of the 3
Surrogate Standards in this sample was above its acceptance range. The
breakdown for Endrin in the Breakdown Standard was 68%. The injection port is
kept at 280 degrees Celcius which may be contributing to high Endrin Breakdown.
An investigation into modifying injection port parameters will be made.