ICPMS daily performance check report, May 14, 2002

Machine-extracted title · confidence 98%

Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 14, 2002.

NYC-WTC_000053389–000053503

Folder label: “ICPMS 5/14/02

Page image
Scanned page image, NYC-WTC_000053443
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document

Ni 60 96.820 Cu 63 94.712 Cu 65 95.355 Zn 66 99.959 Zn 67 100.458 Zn 68 100.370 Ge 72 84.053 As 75 103.398 Se 77 109.501 Se 82 109.320 Kr 83 Ag 107 90.898 in 115 85.098 Sb 121 109.278 Sb 123 110.586 Ba 135 90.713 Ba 137 88.742 > Tb 159 97.828 Hg 201 94.068 Hg 202 93.323 TI 205 93.659 Pb 208 88.115 Calibration Analyte MassCurve Type Correlation Coefficient Be 9Linear Thru Zero 0.999981 Sc-1 45Linear Thru Zero Cr 52Linear Thru Zero 0.999904 Cr 53Linear Thru Zero 0.999990 Mn 55Linear Thru Zero 0.999989 Ni 60Linear Thru Zero 0.999999 Cu 63Linear Thru Zero 0.999999 Cu 65Linear Thru Zero 0.999996 Zn 66Linear Thru Zero 0.999990 Zn 67Linear Thru Zero 0.999992 Zn 68Linear Thru Zero 0.999998 > Ge 72Linear Thru Zero As 75Linear Thru Zero 0.999996 Se 77LlnearThru Zero 0.999964 Se 82Linear Thru Zero 0.999994 Kr 83Linear Thru Zero Ag 107Linear Thru Zero 0.999985 > In 115Linear Thru Zero Sb 121 Linear Thru Zero 0.999937 Sb 123Linear Thru Zero 0.999881 Ba 135Linear Thru Zero 0.999906 Ba 137Linear Thru Zero 0.999955 . Tb 159Linear Thru Zero Hg 201Linear Thru Zero 0.999776 Hg 202Linear Thru Zero 0.999858 TI 205Linear Thru Zero 0.999484 Pb 208Linear Thru Zero 0.999961

Sample ID: AE09766 Report Date/Time: Tuesday, May 14, 2002 10:58:16 Page 2

NYC-WTC_000053443

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000053443Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Page similarity is temporarily unavailable or this page has no indexed vector.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →