NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
NYC-WTC_000040976–000041063Folder label: “ICPMS 4/23/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document
WESTCHESTER COUNTY DEPT OF LABS AND RESEARCH Method 200.8 - Summary Report Sample ID: CCV STD2 60 PPB AND 5 PPB Sample Date/Time: Tuesday, April 23, 2002 11:50:36 Sample Type: Sample Sample Description: Number of Replicates: 3 Method File: d:\elandata\Method\NYC.mth Dataset File: d:\elandata\Dataset\new york city\CCV STD2 60 PPB AND 5 PPB.681 Aliquot Volume (mL): Diluted To Volume (mL):
Concentration Results Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit Be 9 23532.326 1.813 0.049 65.47491 J 0.33 ug/L > Se-1 45 484421.644 1.714 484421.644 ug/L Cr 52 529383.720 2.161 1.077 59.04943 1.21 ug/L Cr 53 62875.692 1.807 0.128 59.16870 1.27 ug/L Mn 55 832201.706 2.959 1.716 60.61172 1.97 ug/L Ni 60 128294.849 2.286 0.264 58.72238 0.62 ug/L Cu 63 288546.059 0.801 0.595 58.07776 1.50 ug/L Cu 65 143033.487 2.009 0.295 57.64908 1.44 ug/L Zn 66 93767.201 2.334 1.009 60.67241 1.87 ug/L Zn 67 16913.890 0.964 0.182 59.76650 0.88 ug/L Zn 68 69758.373 1.812 0.751 60.34070 2.12 ug/L Ge 72 92112.271 1.282 92112.271 ug/L As 75 107985.353 1.995 1.172 59.73505 2.37 ug/L Se 77 10010.334 1.217 0.107 60.38625 2.44 ug/L Se 82 11899.235 0.671 0.128 59.25250 0.62 ug/L Ag 107 560971.780 2.055 0.447 58.97160 2.75 ug/L Cd 111 149538.950 1.171 0.119 59.84012 1.79 ug/L Cd 114 354043.674 0.866 0.283 60.17661 2.25 ug/L In 115 1252981.421 1.751 1252981.421 ug/L Sb 121 502706.186 1.254 0.401 59.00801 1.27 ug/L Sb 123 391371.946 1.305 0.312 59.28262 0.45 ug/L Ba 135 149872.172 0.868 0.087 59.43707 0.91 ug/L Ba 137 261857.283 1.061 0.152 59.30130 0.89 ug/L > Tb 159 1720912.913 0.307 1720912.913 ug/L Ho 165 1705103.072 0.975 -0.001 ug/L Hg 201 5491.380 0.919 0.003 5.01140 0.74 ug/L Hg 202 12305.781 1.788 0.007 4.81207 1.87 ug/L n 205 1358722.123 1.115 0.789 58.83643 1.20 ug/L Pb 208 1835418.971 0.638 1.065 59.23736 0.87 ug/L
QC Calculated Values Analyte Mass QC Std % Recovery Int Std % Recovery Spike % Recovery Dilution % Diff Dup. Rei. % Diff ' Be 9 > Se-1 45 104.321 Cr 52
Sample ID: CCV STD2 60 PPB AND 5 PPB Report Date/Time: Tuesday, April 23, 2002 11:53:05 Page 1
NYC-WTC_000041061
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
No available related records indexed in this group.
NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077433
Summary of water sample results for metals analysis using Method 200.8 conducted on March 21, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077357
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077278
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077205
Summary of water sample results for metals analysis using Method 200.8 conducted on March 15, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000076792
Summary of water sample results for metals analysis using Method 200.8 conducted on March 6, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053212
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 9, 2002.
1 shared topics · similarity 0.990
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077560
Laboratory report from Westchester County Dept of Labs and Research detailing Method 200.8 metal concentrations in water samples dated March 21, 2002.
1 shared topics · similarity 0.987
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053630
Laboratory analysis results for a Westchester County Department of Labs and Research sample processed on May 15, 2002.
1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053505
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 15, 2002.
1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054375
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 24, 2002.
1 shared topics · similarity 0.985
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
1 shared topics · similarity 0.985
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
Page similarity is temporarily unavailable or this page has no indexed vector.
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000040975
Document containing ICPMS analytical data for the World Trade Center site from April 2002.
Filed in the same folder
NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 21, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar indexed content