NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
NYC-WTC_000040976–000041063Folder label: “ICPMS 4/23/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document
WESTCHESTER COUNTY DEPT OF LABS AND RESEARCH Method 200.8 - Summary Report Sample ID: ICV CALBK Sample Date/Time: Tuesday, April 23, 2002 09:24:53 Sample Type: Sample Sample Description: Number of Replicates: 3 Method File: d:\elandata\Method\NYC.mth Dataset File: d:\elandata\Dataset\new york cit\ICV CAL BK.656 Aliquot Volume (mL): Diluted To Volume (mL):
Concentration Results Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit Be 9 9.000 44.444 -0.000 -0.00039 2928.73 ug/L Se-1 45 472439.562 2.270 472439.562 ug/L Cr 52 7058.143 1.032 -0.001 -0.04391 23.31 ug/L Cr 53 665.698 4.742 -0.000 -0.07284 22.43 ug/L Mn 55 865.383 9.028 -0.000 -0.00899 52.43 ug/L Ni 60 210.338 9.321 -0.000 -0.00078 1429.53 ug/L Cu 63 415.014 5.060 -0.000 -0.00818 33.27 ug/L Cu 65 220.005 7.144 -0.000 -0.01002 58.33 ug/L Zn 66 712.369 17.200 -0.001 -0.03517 235.55 ug/L Zn 67 138.336 7.730 -0.000 -0.10534 37.92 ug/L Zn 68 547.689 16.935 -0.000 -0.03881 216.10 ug/L > Ge 72 90186.708 0.236 90186.708 ug/L As 75 63.726 67.927 0.000 0.00685 356.49 ug/L Se 77 109.217 19.264 -0.000 -0.00358 3665.66 ug/L Se 82 104.002 7.510 -0.000 -0.00360 1083.18 ug/L Ag 107 2397.010 12.549 0.001 0.17767 16.00 ug/L Cd 111 984.585 4.424 -0.000 -0.02763 73.72 ug/L Cd 114 125.321 16.559 -0.000 -0.00460 77.83 ug/L > In 115 1302315.641 0.905 1302315.641 ug/L Sb 121 55.334 13.564 0.000 0.00077 103.06 ug/L Sb 123 51.656 25.628 0.000 0.00174 106.70 ug/L Ba 135 79.334 5.684 -0.000 -0.00245 83.74 ug/L Ba 137 132.669 1.569 -0.000 -0.00035 228.21 ug/L > Tb 159 1824618.367 1.505 1824618.367 ug/L Ho 165 1805303.765 1.274 -0.002 ug/L Hg 201 104.335 13.868 0.000 0.04061 27.41 ug/L Hg 202 210.338 7.582 0.000 0.02599 18.35 ug/L TI 205 756.372 2.568 0.000 0.00236 21.98 ug/L Pb 208 1806.421 2.773 -0.000 -0.00428 /16.55 ug/L
QC Calculated Values Analyte Mass QC Std % Recovery Int Std % Recovery Spike % Recovery Dilution % Diff Dup. Rei. % Diff Be 9 > Sc-1 45 101.740 Cr 52
Sample ID: ICV CALBK Report Date/Time: Tuesday, April 23, 2002 09:27:25 Page 1
NYC-WTC_000040986
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
No available related records indexed in this group.
NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077433
Summary of water sample results for metals analysis using Method 200.8 conducted on March 21, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077357
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077278
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077205
Summary of water sample results for metals analysis using Method 200.8 conducted on March 15, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000076792
Summary of water sample results for metals analysis using Method 200.8 conducted on March 6, 2002.
1 shared topics · similarity 0.991
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053212
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 9, 2002.
1 shared topics · similarity 0.990
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077560
Laboratory report from Westchester County Dept of Labs and Research detailing Method 200.8 metal concentrations in water samples dated March 21, 2002.
1 shared topics · similarity 0.987
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053630
Laboratory analysis results for a Westchester County Department of Labs and Research sample processed on May 15, 2002.
1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053505
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 15, 2002.
1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000054375
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 24, 2002.
1 shared topics · similarity 0.985
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053939
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 21, 2002.
1 shared topics · similarity 0.985
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
Page similarity is temporarily unavailable or this page has no indexed vector.
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000040975
Document containing ICPMS analytical data for the World Trade Center site from April 2002.
Filed in the same folder
NYC-WTC_000052963
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 3, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 21, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar indexed content
Summary of water sample results for metals analysis using Method 200.8 conducted on March 19, 2002.
Similar indexed content