NYC-WTC_000006403
Calibration table listing retention times and responses for deuterated standards including naphthalene-d8.
1 shared topics · Same agency, volume and box · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗Initial GCMS analysis report for sample AE01940 conducted by analyst DLV on March 13, 2002.
NYC-WTC_000006371–000006376Folder label: “GCMS X 2/15/02 A 2/9/02 & 2/16/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document gUcLIL. tatl ACPOL L IC V L wcu)
Data File C:\HPCHEM\l\DATA\FEB0802\1940.D Vial: 20 Acq On 9 Feb 2002 3:48 am Operator: 209 GALOS Sample GC/MS SCAN 1/28 Inst : GC/MS 209 Mise Multiplr: 1.00 MS Integration Params: GOOFY.P Quant Time: Feb 13 8:40 2002 Quant Results File: GCMS0208.RES
Quant Method : C:\HPCHEM\1\METHODS\GCMS0208.M (RTE Integrator) Title : 209 625/TCLP calibration table Last Update : Fri Feb 08 15:29:22 2002 Response via : Initial Calibration DataAcq Meth : GCMS0208
Internal Standards R.T. Qlon Response Cone Units Dev(Min)
1) 1,4-Dichlorobenzene-d4 12.35 152 321371 20.00 ug/1 0.00 10) Naphthalene-d8 15.99 136 1221138 20.00 ug/1 0.00 17) Acenaphthene-diO 21.31 164 648401 20.00 ug/1 0.00 29) Phenanthrene-diO 25.76 188 934553 20.00 ug/1 0.00 38) Chrysene-dl2 33.83 240 634504 20.00 ug/1 -0.02 44) Perylene-dl2 38.12 264 431233 20.00 ug/1 0.00
System Monitoring Compounds 37) 2,4-DDD 30.84 235 39358 3.39 0.34 Spiked Amount 5.000 Recovery 67.80%
Target Compounds Qvalue 2) N-nitroso-dimethyl amine 0.00 74 0 N.D. 3) bis(2-Chloroethyl)ether 0.00 93 0 N.D. 4) 1,3-Dichlorobenzene 0.00 146 0 N.D. 5) 1,4-Dichlorobenzene 0.00 146 0 N.D. 6) 1,2-Dichlorobenzene 0.00 146 0 N.D. 7) 2,21-oxybis(1-Chloropropan 0.00 45 0 N.D. 8) N-Nitroso-di-n-propylamine 0.00 70 0 N.D. 9) Hexachloroethane 0.00 117 0 N.D. 11) Nitrobenzene 0.00 77 0 N.D. 12) Isopho rone 0.00 82 0 N.D. 13) bis(2-Chloroethoxy)methane 0.00 93 0 N.D. 14) 1,2,4-Trichlorobenzene 0.00 180 0 N.D. 15) Naphthalene 16.05 128 611 N.D. 16) Hexachlorobutadiene 0.00 225 0 N.D. 18) Hexachlorocyclopentadiene 0.00 237 0 N.D. 19) 2-Chloronaphthalene 0.00 162 0 N.D. 20) Dimethylphthalate 0.00 163 0 N.D. 21) 2,6-Dinitrotoluene 0.00 165 0 N.D. 22) Acenaphthylene 0.00 152 0 N.D. 23) Acenaphthene 0.00 153 0 N.D. 24) 2,4-Dinitrotoluene 0.00 165 0 N.D. 25) Diethylphthalate 0.00 149 0 N.D. 26) 4-Chlorophenyl-phenylether 0.00 204 0 N.D. 27) Fluorene 0.00 166 0 N.D. 28) Azobenzen/ 1,2-Diphenylhyd 0.00 77 0 N.D.
(#) = qualifier out of range (m) = manual integration 1940.D GCMS0208.M Fri Mar 08 15:28:22 2002 Page 1
NYC-WTC_000006371
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
NYC-WTC_000006403
Calibration table listing retention times and responses for deuterated standards including naphthalene-d8.
1 shared topics · Same agency, volume and box · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000006387
Calibration table listing retention times and responses for deuterated standards including naphthalene-d8.
1 shared topics · Same agency, volume and box · similarity 0.995
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052600
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
1 shared topics · similarity 0.995
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052666
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
1 shared topics · similarity 0.994
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052614
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
1 shared topics · similarity 0.993
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052736
GC/MS analysis report for sample AE01037 showing surrogate recovery data generated on February 26, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000009686
Laboratory file showing GC/MS calibration response factors for surrogate compounds from February and March 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052714
Calibration table and instrument data for GC/MS analysis generated on February 12 and 19, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052586
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052701
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000009670
Laboratory file showing GC/MS calibration response factors for surrogate compounds from February and March 2002.
1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052630
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
Similar document embedding; no shared entity or topic indexed. · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
NYC-WTC_000006387
Calibration table listing retention times and responses for deuterated standards including naphthalene-d8.
Similarity score 0.994 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052714
Calibration table and instrument data for GC/MS analysis generated on February 12 and 19, 2002.
Similarity score 0.993 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052630
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
Similarity score 0.993 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052918
Laboratory report documenting gas chromatography-mass spectrometry analysis of sample AE01402 conducted on February 27, 2002.
Similarity score 0.993 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052693
Laboratory report documenting GCMS analysis of sample AE00945 conducted on February 26, 2002.
Similarity score 0.993 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000052701
Calibration record for GC/MS instrument 209 used in TCLP testing, listing standard compounds and response factors from February 14, 2002.
Similarity score 0.993 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000006359
Department of Environmental Protection GCMS sample tracking record for the World Trade Center site dated February 2002.
Filed in the same folder
Header page for a GCMS log sheet covering the period from February 15 to March 2, 2002.
Filed in the same folder
NYC-WTC_000006362
Initial GCMS analysis report for sample AE01940 conducted by analyst DLV on March 13, 2002.
Filed in the same folder
NYC-WTC_000006377
Quality control page documenting surrogate compound recovery rates for a gas chromatography mass spectrometry run.
Filed in the same folder
NYC-WTC_000006387
Calibration table listing retention times and responses for deuterated standards including naphthalene-d8.
Filed in the same folder