Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - September 29, 2005
8.2 Measurement Performance Criteria The 133-135 Greenwich Street and 21-23
Thames Street Ambient Air Monitoring Program is designed to produce data of
the quality necessary to achieve PQOs and meet or exceed the minimum standard
requirements for field and analytical methods. The overall QA objective is to develop and
implement procedures for field sampling, chain-of-custody, laboratory analysis, and
reporting which will provide results that are scientifically valid, and the levels of which
are sufficient to meet PQOs. Specific procedures for sampling, chain of custody,
laboratory and field instruments calibration, laboratory analysis, reporting of data,
internal quality control, preventative maintenance of field and laboratory equipment, and
corrective action are described in other sections of this QAPP. The purpose of this section
is to state the specific, required QA objectives for accuracy, precision, representativeness,
completeness, sensitivity, selectivity, and comparability. Measurement performance
criteria for precision, accuracy/bias, representativeness, completeness, sensitivity,
quantitation limits, selectivity, and comparability have been established for each
parameter and are summarized in Tables 8-2 through 8-9. These measures of
performance are also referred to as Data Quality Indicators (DQIs) and are discussed in
detail below.
8.2.1 Precision
Precision is the agreement among a set of replicate measurements without consideration
of the “true” or accurate value: i.e., variability between measurements of the same
material for the same analyte. Precision is measured in a variety of ways including
statistically, such as calculating variance or standard deviation. The results of the
background phase will be used to monitor overall precision (field and laboratory) and
will be included as Attachment B to the QAPP, when available.
Field Precision Objectives Field precision is assessed through the collection and
measurement of collocated samples (also called field duplicates) which consist of a
second sample in addition to the original field sample. In general, field duplicates will be
collected at a frequency of once per week per analytical parameter. Precision will be
measured through the calculation of relative percent difference (RPD). The resulting
information will be used to assess sample homogeneity, spatial variability at the site,
sample collection reproducibility, and analytical variability. Field duplicate RPDs must
be < 40. Field precision will be maintained by utilizing experienced/trained
sampling crews and conducting field audits. In addition, during the background phase of
the program, the precision of the mercury sampling and analysis method will be tested.
Ambient air samples will be collected at one location using a series of four iodated
carbon traps prespiked (50 ng/trap) with mercury. These samples will provide data on the
precision of the method under actual field sampling conditions. Following the
background phase, one field spike will be performed every other week, alternating with
the field duplicates.
Laboratory Precision Objectives Precision in the laboratory is assessed through the