NYC 9/11 Public Portal Document
Part V - Quality Assurance Project Plan
133-135 Greenwich Slreet/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
The maintenance responsibilities for fixed laboratory instruments will be assigned to
the Laboratory Section Managers. Laboratory analysts will be responsible for daily
checks and calibrations and for reporting any problems with the instruments. The
maintenance schedule will follow the manufacturer’s recommendations. Laboratory
personnel will also be responsible for ensuring that critical parts are kept with the fixed
laboratory instruments. Critical spare parts will be immediately available to reduce
potential downtime. The inventory will primarily contain parts that are subject to frequent
failure, have limited useful lifetimes, and/or cannot be obtained in a timely
manner. Annual preventative maintenance service visits will involve cleaning, adjusting,
inspecting, and testing procedures designed to minimize product failure and/or extend the
product’s life. Between visits, laboratory analysts will be responsible for performing
routine operator maintenance and cleaning in accordance with the manufacturer’s
specifications.
13.2 Instrument/Equipment Calibration and Frequency
13.2.1 Field Equipment
All materials, including standards or standard solutions, will be dated upon receipt, and
will be identified by material name, lot number, purity or concentration, supplier,
recipient’s name, and expiration date. All materials must be National Institute of Standard
and Technology (NIST)- traceable reference materials.
Metals (Low-Volume Area Sampler) Table 13-3a summarizes the calibration
procedures associated with the low volume samplers used to collect metals (MCE)
samples._______________________________________________________________
Table 13-3a. Calibration Requirements for Low Volume Area Samplers
Equipment Activity Acceptance Corrective Action
Criteria
Low Volume Area Rotometer calibration at Within Method Replace Sampler
Sampler every sample run (NIOSH 7300)
initiation and accepted range
termination. Periodic 1-4 1pm
check during run.
Calibration records maintained and tracked by Quality Assurance Officer and Field Sampling
Coordinator.
91
CITY_21MC100_007120943
NYC-WTC_000443832
OCR can misread numbers and units. Confirm readings against the page image before using them.