Part V - Quality Assurance Project Plan
133-135 Greenwich Slreet/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
project-specific target analytical sensitivity of less than 0.0009 libers cmt. In addition,
samples may also be analyzed using Scanning Electron Microscopy (SEM) due to the
potential overloading problem with TEM which may be caused by excess particulates.
This will be determined on a sample-specific basis and performed as needed. SEM is a
well-established methodology that utilizes a scanning electron microscope. The SEM can
work in many different magnification ranges from lOOx to 20,000x. The sample
preparation is different than TEM, whereas samples are gold coated and mounted on a
stub, with minimal degradation to the original sample. The SEM is more flexible with
regard to types of material you can introduce into the chamber. SEM can achieve much
better contrast than TEM and results in almost a 3-D image whereas the TEM is a 2-D
“shadow”. SEMs are typically equipped with Energy Dispersive X-Ray analyzers
(EDXA) just like TEM. This enables the analyst to determine the elemental composition
of a material or fiber observed. The SEM is not equipped with Electron Diffraction
capabilities unlike the TEM, which allows the analyst to determine the crystal structure of
a mineral based on a distinct pattern of spots that is produced from the bombardment of
an electron beam. Selected Area Electron Diffraction (SAED) is cited in many TEM
methods including AHERA and is required to yield a definitive asbestos
identification. For this project, however, since the dust has been so well characterized, the
need for SAED is not as essential.
11.1.2.4 Metals Analysis by SW-846 Method 6020A
Metals analysis will be performed using SW-846 Method 6020A with the
following modifications:
• Samples will be digested using nitric acid as per 40 CFR Part 50 Appendix J, not
a mixture of nitric and hydrochloric acid. Sample digestion will be performed
using a hotplate or Modblock digestion procedure.
• Initial ICP/MS instrument calibration will be performed using a blank and a
single-point standard.
• The routine calibration procedure will not include the analysis of a High
Standard Verification (HSV) standard. Calibration will be verified with the CCB
and CCV.
• The laboratory will perform the metals analysis by scanning the selected ion, not
scanning all ions from 5-250 amu.
• The internal standard mixture used by the laboratory will include Li, Ge, In, Tm,
and Bi.
11.1.2.5 PCB Analysis
PCB extracts for Aroclor analysis will be concentrated to a final volume of 5 mL. If the
total PCB analysis yields a result which exceeds the USEPA Site-Specific Trigger Level,
a further congener analysis will be performed on the same extract to confirm the total
PCB concentration. This analysis will also be performed using GC/ECD; however,
further QC analyses for the PCB congener analysis will not be required assuming the
72
CITY_21MC100_007120924
NYC-WTC_000443813