NYC 9/11 Public Portal Document
Part V - Quality Assurance Project Plan
133-135 Greenwich Streel/21-23 Thames Street
Airtek Project 05-0701 - Rev 1 - January 27, 2006
Metals (Low -Volume Area Sampler)
Table 13-la summaiizes the inspection, testing, and maintenance activities associated
with the high volume samplers used to collect metals samples.
Table 13-la. Maintenance, Testing, and Inspection Activities Associated with
Metals (MCE filters) Collected with Low Volume Area Samplers
Eau/cment Activity Acceptance Corrective Action
Criteria
Power Cords Check for crimps or No obvious damage Replace as
cracks necessary
Motor Replace as needed Consult Obtain the correct
manufacturer for model
correct model of
motor
Tubing and fittings Visually inspect No crimps, cracks, Replace as
daily or obstructions; no necessary
crossthreading
Mercury (Gaseous) Using a Direct Read Analyzer
Table 13-lb summarizes the inspection, testing, and maintenance activities associated
with the gaseous mercury analysis using the Ohio Lumex R A91 5-h mercury analyzer._______________
Table 13-lb. Maintenance, Testing, and Inspection Requirements for the Ohio Lumex RA9I5 +
MercurjAiialyzcr
Fault Symptom Possible Cause Corrective Action
Segments of the indicator -Power cable is out of order -Repair the power cable
table on the display and
control unit are not -Display unit cable is out of -Repair the display unit
highlighted when the order
analyzer is switched on.
-Battery is discharged -Charge the battery
88
CITY_DEP_WORBY_CANNATA_SUBP_000093276
NYC-WTC_000427044
OCR can misread numbers and units. Confirm readings against the page image before using them.