Quality Control for PCDD/PCDF Analyses
The performance of the sample processing steps and the instrumentation are monitored on a
routine basis. The procedures and criteria are summarized below.
One method blank and one laboratory spike sample are typically prepared with each ten samples of
any given matrix. Recoveries of the native PCDD/PCDF analytes in the laboratory spike samples
generally range from 70 to 130%. Recoveries of selected analytes outside this range do not
invalidate the data but provide information, which is used by the laboratory to monitor recovery
trends and to assure optimization of the method.
" Internal standards are spiked into each sample prior to extraction in order to monitor the level of
recovery, which is achieved for each individual sample. Acceptable recoveries range from 40 to
i 135 percent for the internal standards unless a deviation is due to variation in instrument response
as a result of analytical interferences.
: The resolution of the mass spectrometer is verified prior to each analysis to be 10,000 or greater.
Hardcopies of the reference peaks are printed at the beginning and end of each analysis day. The
s resolving power of the DB-5MS chromatographic column is checked daily by analyzing a standard
3 solution containing 2,3,7,8-TCDD and the adjacent TCDD isomers. The DB-2.25 column resolution
is checked daily by analyzing a standard solution containing 2,3,7,8-TCDF and the adjacent TCDF
4 isomers. Acceptable performance is achieved when 2,3,7,8-TCDD or 2,3,7,8-TCDF is resolved
y from the adjacent isomers by a valley of 25% or less. The group times for the selected-ion
monitoring data acquisitions are also checked daily by analyzing the column performance mix
which has been modified to contain the first and last eluting isomers of each congener class. In
this way one is assured of collecting data representative of the total PCDD/PCDF content and that
the 2,3,7,8-substituted isomers are suitably resolved.
u Initial calibrations are generated by analyzing standard solutions (see Table 4] containing target
native and labeled PCDD/PCDF compounds. Response factors are calculated and averaged for
3 each compound. These averages are used for quantification and for comparison to the daily
ri continuing calibration. The relative standard deviation for each native compound must be 20% or
less (30% or less for the labeled compounds) as specified in Method 8290. A continuing
calibration standard is analyzed at the beginning and end of each 12-hour shift on days when initial
calibrations are not performed. The initial calibration is considered to be valid when the response
factors from the continuing calibration analysis fall to within the ranges specified in Method 8290.
REPORT OF LABORATORY ANALYSIS
This report shall not be reproduced, except in full,
without the written consent of Pace Analytical Services, Inc.