Pace Analytical Senfices, Inc. 1700 Elm Street, Suite 200 ‘ace Analytical' > www.pacelabs.com REPORT OF; CHEMICAL ANALYSES Minneapolis, MN 55414 Phone: 612.607.1700 . Fax: 612.607.6444
PROJECT; PCDD/PCDF ANALYSES DATE: January 9, 2002 • PAGE: 6 REPORT NO: 01-1052682 Cm < 'Y B Quality Control for PCDD/PCDF Analyses
The performance of the sample processing steps and the instrumentation are monitored on a routine basisi"The procedures and criteria are summarized below.
One m'efhod blank and one laboratory spike sample are typically prepared with each ten samples of any; given matrix. Recoveries of the native PCDD/PCDF analytes in the laboratory spike samples R.: vgenerally range from 70 to 130%. Recoveries of selected analytes outside this range do not invalidate the data but provide information, which is used by the laboratory to monitor recovery trends and to assure optimization of the method.
Internal standards are spiked into each sample prior to extraction in order to monitor the level of 12 recovery, which is achieved for each individual sample. Acceptable recoveries range from 40 to 135 percent for the internal standards unless a deviation is due to variation in instrument response f. as a result of analytical interferences.
The resolution of the mass spectrometer is verified prior to each analysis to be 10,000 or greater. Hardcopies of the reference peaks are printed at the beginning and end of each analysis day. The resolving power of the DB-5MS chromatographic column is checked daily by analyzing a standard solution containing 2,3,7,8-TCDD and the adjacent TCDD isomers. The DB-225 column resolution . is checked daily by analyzing a standard solution containing 2,3,7,8-TCDF and the adjacent TCDF isomers. Acceptable performance is achieved when 2,3,7,8-TCDD or 2,3,7,8-TCDF is resolved from the adjacent isomers by a valley of 25% or less. The group times for the selected-ion- - monitoring data acquisitions are also checked daily by analyzing the column performance mix I which has been modified to contain the first and last eluting isomers of each congener class. In J this way one is assured of collecting data representative of the total PCDD/PCDF content and that the 2,3,7,8-substituted isomers are suitably resolved.
LJ Initial calibrations are generated by analyzing standard solutions (see Table 4) containing target native and labeled PCDD/PCDF compounds. Response factors are calculated and averaged for Q each compound. These averages are used for quantification and for comparison to the daily continuing calibration. The relative standard deviation for each native compound must be 20% or less (30% or less for the labeled compounds) as specified in Method 8290. A continuing I calibration standard is analyzed at the beginning and end of each 12-hour shift on days when initial J calibrations are not performed. The initial calibration is considered to be valid when the response factors from the continuing calibration analysis fall to within the ranges specified in Method 8290.
REPORT OF LABORATORY ANALYSIS This report shall not be reproduced, except in full, without the written consent of Pace Analytical Services, Inc.
CITY_21MC1OO_OO2896256
NYC-WTC_000226909
OCR can misread numbers and units. Confirm readings against the page image before using them.
Related records
Browse subjects →Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
Same-box records
9 shared non-person entities · 1 shared topics · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗9 shared non-person entities · 1 shared topics · similarity 0.992
machine-extracted · confidence unavailable · Check page ↗8 shared non-person entities · 1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000175725
Confidential attachment containing analytical laboratory reports regarding polychlorinated dibenzo-p-dioxins and furans under a protective order.
9 shared non-person entities · similarity 0.982
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000175186
Cover page for a laboratory report containing 17 pages of analytical results conforming to NELAC standards for the Department of Education.
5 shared non-person entities · 1 shared topics · similarity 0.974
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000178033
Confidential laboratory analysis report appendix from [name] Services regarding World Trade Center samples.
3 shared non-person entities · 1 shared topics · similarity 0.972
machine-extracted · confidence unavailable · Check page ↗3 shared non-person entities · 1 shared topics · similarity 0.969
machine-extracted · confidence unavailable · Check page ↗3 shared non-person entities · 1 shared topics · similarity 0.968
machine-extracted · confidence unavailable · Check page ↗3 shared non-person entities · 1 shared topics · similarity 0.966
machine-extracted · confidence unavailable · Check page ↗3 shared non-person entities · 1 shared topics · similarity 0.966
machine-extracted · confidence unavailable · Check page ↗3 shared non-person entities · 1 shared topics · similarity 0.964
machine-extracted · confidence unavailable · Check page ↗Filed elsewhere
No available related records indexed in this group.
More like this page
Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
Similarity score 0.988 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Similarity score 0.988 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Similarity score 0.987 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Similarity score 0.976 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Similarity score 0.952 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Similarity score 0.951 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Others also read
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
Similar indexed content
Similar indexed content
Similar indexed content
Similar indexed content
NYC-WTC_000175725
Confidential attachment containing analytical laboratory reports regarding polychlorinated dibenzo-p-dioxins and furans under a protective order.
Similar indexed content