NYC 9/11 Public Portal Document I Search - 1 Result - APPENDIX A TO SUBPART E - INTERIM TRANSMISSION EL... Page 30 of 38

magnification (15,000 to 20,000X screen magnification); and analyze until the stopping rule becomes applicable.

q. Carefully record all data as they are being collected, and check for accuracy.

r. After finishing with a grid, remove it from the microscope, and replace it in the appropriate grid hold. Sample grids must be stored for a minimum of 1 year from the date of the analysis; the sample cassette must be retained for a minimum of 30 days by the laboratory or returned at the client's request.

H. Sample Analytical Sequence

1. Carry out visual inspection of work site prior to air monitoring.

2. Collect a minimum of five air samples inside the work site and five samples outside the work site. The indoor and outdoor samples shall be taken during the same time period.

3. Analyze the abatement area samples according to this protocol. The analysis must meet the 0.005 s <3> analytical sensitivity.

4. Remaining steps in the analytical sequence are contained in Unit IV. of this Appendix.

I. Reporting

The following information must be reported to the client. See the following Table II:

♦ Display Image

1. Concentration in structures per square millimeter and structures per cubic per cubic centimeter.

2. Analytical sensitivity used for the analysis.

3. Number of asbestos structures.

4. Area analyzed.

5. Volume of air samples (which was initially provided by client).

6. Average grid size opening.

7. Number of grids analyzed.

8. Copy of the count sheet must be included with the report.

9. Signature of laboratory official to indicate that the laboratory met specifications of the AHERA method.

10. Report form must contain official laboratory identification (e.g., letterhead).

11. Type of asbestos.

J. Calibration Methodology

http://lexis/research/retrieve ? m=el35£2ca3d266867e53d261df3205722&_fhitstr=FULL(S... 02/04/2002 CITY_21MC100_012558694

NYC-WTC_000219954

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000219954Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

No available related records indexed in this group.

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →