NYC 9/11 Public Portal Document
I ATC Environmental, Inc.
104 East 25th Street
New York, New York 1001C
ATC Associates Inc. Approved N.Y. State ELAP #10879
104 East 25th Street ACCREDITED NVLAP Lab Code # 1187
New York, NY 10010 Report Date: 10/26/01
ASBESTOS FIBER ANALYSIS BY TRANSMISSION ELECTRON MICROSCOPY (TEM)
With Fiber Identification by Selected-Area Electron Diffraction (SAED) and Energy-Dispersive X-RAY Spectroscopy (EDS;
Method: AHERA Final Rule, EPA Federal Register (40 CFR, Part 763). Vol. 52, Nc. 210, October 30, 1987
Sample Identification Analysis Parameters
Project/Site: Tribeca Towers, 105 Duane St. , NY, NY TEM Sample No.: 5070 - 62561
Hallways
Date Received: 10/26/01 Grid Openings Analyzed: 6
Project Sample No.: H-7 Average Grid Opening: 106.00 um
Air Volume: 1300 liters Total Area Analyzed: 0.0674 mm2
Filter Type: MCE 25 mm diem., Pore Size: 0.45 um Analytical Sensitivity: 14.8 5/mmZ
Effective Area: 385 mm2 0.0C44 s;'cc
-----------------------------------------------'----------------------------------------------------------------------------------
FIBROUS PARTICULATES ANALYZED
Number Observed CONCENTRATION
I Single
Fibers Bundles Clusters
Matrix
Struct.
Total.
Struct.
Total
Struct.
>=5um
Filter
Struct/mm2 Struct/cc
Air
norm/m3
Chrysotile Asbestos 0 0 0 0 0 0 0.0 0.0000 0
Amphibole Asbestos 0 0 0 0 0 0 0.0 0.0000 0
Ambiguous 0 0 0 0 0 0 0.0 0.0000 0
Total asbestos (includes ambiguous structures) 0 0 0.0 0.0000 0
Non-Asbestos 0 0 0 0 0 0 0.0 0.0000 N/A
No Identification 0 0 0 0 0 0 0.0 0.0000 N/A
Totals (All Structures) 0 0 0.0 0.0000 N/A
The estimated uncertainty of this analysis is 1 standard deviation.
----------------------------------------------------------------------------------------------------------------------------------
' Non-asbestos fiber i.d. (EDS/SAED): None detected
Amphibole Asbestos Type: None detected
-----------------------------------------------------------
' The Analysis was carried out in accordance wi:h the approved TEM method.
This laboratory is in compliance with the quality specified for the method.
' Analyzed by: Report Reviewed by:
Marek Peysakhov Michael A. Gorycki, Ph.D.
Director of Electron Microscopy
I
NYC-WTC 000104330