DEP Method 200.8 metal analysis results, Jan 7, 2002

Machine-extracted title · confidence 100%

Lab report detailing ICP-MS analysis of water samples for metals using Method 200.8 on January 7, 2002.

NYC-WTC_000086174–000086255

Folder label: “NEWQ 01/07/2002 31249 - 31403”

Page image
Scanned page image, NYC-WTC_000086254
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document

Method 200.8 - Summary Report Sample ID: CCV STD2 60 PPB Sample Date/Time: Monday, January 07, 2002 16:23:20 Sample Type: Sample Sample Description: Number of Replicates: 3 Batch ID: Method File: d:\elandata\Method\NEWQ200.8.mth Dataset File: d:\elandata\Dataset\daily performance\CCV STD2 60 PPB.1838 Sample Prep Volume (ml): Initial Sample Quantity (mg): Aliquot Volume (ml): Diluted To Volume (mL):

Concentration Results Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit r Be 9 30250.567 2.322 0.045 61.50378 1.33 ug/L I: Sc-1 45 668149.083 1.846 668149.083 ug/L I Cr 52 713623.183 1.781 1.052 58.17124 2.59 ug/L I Cr 53 84603.135 2.057 0.125 57.96717 1.55 ug/L I Mn 55 1119316.950 2.796 1.674 58.11919 4.10 ug/L I Ni 60 178634.011 4.120 0.267 57.83349 5.17 ug/L I Cu 63 394989.137 1.842 0.590 56.89407 3.62 ug/L L Cu 65 193992.577 0.391 0.290 57.45604 1.53 ug/L r Zn 66 130067.521 2.476 1.139 60.43349 4.91 ug/L I Zn 67 22873.983 2.719 0.200 60.05882 0.98 ug/L I Zn 68 96558.594 2.519 0.845 60.53382 4.90 ug/L I: Ge 72 113633.899 2.616 113633.899 ug/L I As 75 134408.593 2.181 1.179 59.93812 4.61 ug/L 1 Se 77 12339.159 1.194 0.107 59.82937 3.78 ug/L L Se 82 15871.355 4.759 0.137 59.99182 7.22 ug/L r Ag 107 649581.157 2.474 0.499 59.68635 3.79 ug/L I: In 115 1302223.862 1.356 1302223.862 ug/L I Sb 121 500848.870 1.706 0.385 60.33901 3.02 ug/L L Sb 123 390649.118 1.443 0.300 60.36747 0.79 ug/L r Ba 135 153593.420 1.772 0.084 58.25385 2.87 ug/L I Ba 137 265624.401 1.716 0.145 58.32159 2.08 ug/L I: Ho 165 1830212.893 2.210 1830212.893 ug/L Tl 205 1726403.189 2.237 0.943 59.10499 2.76 ug/L L Pb 208 2344411.426 1.281 1.279 59.03032 3.12 ug/L

Sample ID: CeV STD2 60 PPB Report Date/Time: Monday, January 07, 2002 16:24:51 Page 1

NYC-WTC_000086254

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000086254Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

ICP-MS summary report, Jan 4 2002 · Box DEP Box 66

NYC-WTC_000085362

Method 200.8 laboratory summary report listing elemental concentrations including Beryllium, Scandium, Chromium, Manganese, Nickel, and Copper.

1 shared topics · Same agency, volume and box · similarity 0.999

machine-extracted · confidence unavailable · Check page ↗
ICP-MS summary report, Jan 3 2002 · Box DEP Box 66

NYC-WTC_000085227

Method 200.8 laboratory summary report listing elemental concentrations including Beryllium, Scandium, Chromium, Manganese, Nickel, and Copper.

1 shared topics · Same agency, volume and box · similarity 0.999

machine-extracted · confidence unavailable · Check page ↗
ICP-MS summary report, Jan 10 2002 · Box DEP Box 66

NYC-WTC_000085591

Method 200.8 laboratory summary report listing elemental concentrations including Beryllium, Scandium, Chromium, Manganese, Nickel, and Copper.

1 shared topics · Same agency, volume and box · similarity 0.998

machine-extracted · confidence unavailable · Check page ↗

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Page similarity is temporarily unavailable or this page has no indexed vector.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →