DESCRIPTION & SPECIFICATIONS
Page 6
SPECIFICATIONS ANALYTICAL-SYSTEM
Dimensions; Analysis- Rate;
W X D - X H (inches)
Sampler 40 X 9 X 10 • Multi-test methods - 120 samples
Analytical Module 20 X 15 X 28 per hour
Data Handler 23 X 24 X 20 • Dedicated Cartridges 120-240 per
Printer 15 X 13 X 14 hour depending on chemistry
Reagent Changer 20 X 15 X 28 selected
Electrical Require-ments: Number of Channels:
Voltage 100, 115, 220, 240 VAC One, two, three or four channels
Frequency - 50/60 Hz depending on configuration selected
Environmental Requirements; Weight (lbs);
Operating Temperature lOOC - 40OC Sampler 55
Relative Humidity up to 95% Analytical Module 120
non-condensing Data Handler 74
Air Pressure minimum 30 PSI - Printer 14
maximum 100 PSI Reagent Cabinet 60
Colorimeter;
Wavelength Range - 340-800 nm
Detector - Solid state Silicon Diode
Absorbance Range - 0.05-3.0 Absorbance Units
Photometric Linearity - .05AU up to l.OAU
Filter Bandwidth - 12-20nm for wavelengths in the 400-800nm region
Flowcell Pathlengths - 10mm, 30mm, 50mm, and 75mm
Photometric &-Electrical Drift - less than 1.5% of full scale per 8 hrs at 0.05AU
Gain Control - Automatically adjusted by computer
Chemical Sensitivity - from less than 1 ug/L to more than 100 g/L depending on
chemistry selected