NYC 9/11 Public Portal Document
Quantitation Report (QT Reviewed)
A
Data File : C:\MSDCHEM\1\DATA\121801\CCD.D Vial: 13
Acq On : 19 Dec 2001 4:54 am Operator: McLean
Sample : 12.5 Inst : Instrumen
Mise Multiplr: 1.00
MS Integration Params: BENZO.P
Quant Time: Dec 19 05:31:05 2001 Quant Results File: 121401.RES
Quant Method : C:\MSDCHEM\l\5973N\121401.M (RTE Integrator)
Title : Quantitation For Method 525.2
Last Update : Tue Dec 18 09:35:41 2001
Response via : Initial Calibration
DataAcq Meth : 121401
Internal Standards R.T. Qlon Response Cone Units Dev(Min)
1) Acenaphthene-diO 18.21 164 8977024 5.00 ug/L 0.00
9) Phenanthrene D-10 22.53 188 12601938 5.00 ug/L 0.00
20) Chrysene D-12 30.31 240 9328772 5.00 ug/L 0.00
System Monitoring Compounds
2) 1,3-Dimethyl-2-nitrobenzen 13.13 134 2236768 4.87 ug/L 0.00
Spiked Amount 5.000 Recovery = 97.40%
19) Triphenyl Phosphate surr 29.58 326 2311835 6.03 ug/L 0.00
Spiked Amount 5.000 Recovery = 120.60%
21) Perylene D-12 surr 34.32 264 6419650 5.20 ug/L 0.00
Spiked Amount 5.000 Recovery = 104.00%
Target Compounds Qvalue
3) Hexachlorocyclopentadiene 15.71 237 3736879 14.70 ug/L 98
4) Hexachlorobenzene 21.41 284 6537105 11.01 ug/L 98
5) EPTC 16.18 128 10931328 12.79 ug/L 99
6) 2,6-Dinitrotoluene 17.78 165 5677404 12.75 ug/L 96
7) 2,4-Dinitotoluene 18.93 165 6788838 12.47 ug/L 98
8) Molinate 19.09 126 15633629 12.91 ug/L 99
10) Simazine 21.98 201 4808330 14.04 ug/L 98
11) Atrazine 22.08 200 6589133 14.72 ug/L 100
12) Alachlor 23.92 160 6067461 13.95 ug/L 93
13) Terbacil 22.94 161 8223960 15.58 ug/L 99
14) Acetochlor 23.71 146 4482108 14.16 ug/L 95
15) Metribuzin 23.76 198 7275440 11.91 ug/L 95
16) Metolachlor 24.84 162 20149351 14.37 ug/L 99
17) Butachlor 26.67 160 5850204 14.46 ug/L 83
18) 4,4'-DDE 27.30 246 5856724 11.93 ug/L 96
22) Bis (2-Ethylhexyl) adipate 29.52 129 18125235 10.91 ug/L 97
23) Bis (2-Ethylhexyl) phthala 30.95 149 22912362m <7.72 uq/L
24) Benzo (a) pyrene 34.17 252 24270307 13.21 ug/L 97
-16
2.19
(#) = qualifier out of range (m) = manual integration
CCD.D 121401.M Thu Dec 20 09:32:18 2001 Page 1
NYC-WTC_000055809
OCR can misread numbers and units. Confirm readings against the page image before using them.